Smiths Medical International Ltd.
Issue 5 (August 2004)
3200 Service Manual
2 — 9
Diagnostic mode
After carrying out the previous display request (i.e. pressing
START
), press
STOP
to return
the pump to its set-up mode.
Transducer
alarm
If a fault occurs within the transducer set-up system, a pulsing loud alarm (silenceable)
occurs and the following alarm display appears:
Battery
3.
voltage
The battery voltage display is similar to the following:
WARNING: CANNOT
CALIBRATE TRANSDUCER
BATTERY VOLTAGE=
6.9 V
Language
4.
Various languages may be selected and used for display purposes. The English display
selection is shown below:
LANGUAGE=
ENGLISH
The other language displays are:
•
SPRÅK
= SVENSKA
•
TAAL
= NEDERLANDS.
•
LINGUA
= ITALIANO.
•
SPRACHE = DEUTSCH.
•
LANGUE = FRANCAIS.
The total number of faults that may have occurred to the pump since manufacture is
displayed when in this parameter:
Number of
5.
faults
Volume
6.
infused
The total volume of liquid that has been infused since the pump was manufactured is
displayed when in this parameter:
NUMBER OF FAULTS=
1
TOTAL VOLUME INFUSED
0.06 litre
Hours of use
7.
The total number of hours that the pump has been switched on is displayed when in this
parameter is selected:
TOTAL HOURS OF USE=
9.0 hours
Summary of Contents for Graseby 3200
Page 15: ...CHAPTER 1 INTRODUCTION FEATURES and SPECIFICATION 3200 In line Pressure Syringe Pump ...
Page 22: ...CHAPTER 2 CONFIGURATION DIAGNOSTICS and OCCLUSION THRUST 3200 In line Pressure Syringe Pump ...
Page 37: ...CHAPTER 3 FUNCTIONAL DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Page 43: ...CHAPTER 4 CIRCUIT DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Page 83: ...CHAPTER 6 FUNCTIONAL TESTS and MANUFACTURNG SETTINGS 3200 In line Pressure Syringe Pump ...
Page 91: ...CHAPTER 7 ILLUSTRATED PARTS LIST 3200 In line Pressure Syringe Pump ...
Page 105: ...CHAPTER 8 BRAUN PERFUSOR CONVERSION 3200 In line Pressure Syringe Pump ...
Page 108: ...CHAPTER 9 DC INPUT VERSION of 3200 3200 In line Pressure Syringe Pump ...
Page 111: ...APPENDIX FITTING a MODIFIED SIZE SENSOR FLAG 3200 In line Pressure Syringe Pump ...