Table 5.1 Main processor fault codes
(contd.)
Fault codes
(contd.)
The five items shown in
Table 5.2
(and cross referenced from
Table 5.1
) detail the
possible cause of a fault within a pump, and the recommended action to be taken:
Item Fault
Possible cause
Recommended action
1
Communication
External interference...
Check communication
(Comms.) fault.
e.g. static interference
link and interface cable.
or r.f. interference.
Circuit fault.
Return to Smiths Medical.
2
Internal fault.
External interference...
Relocate the pump.
e.g. static interference
or r.f. interference.
Circuit fault.
Return to Smiths Medical.
3
Keyboard fault.
Damaged keyboard.
Replace keyboard.
4
Motor fault.
Faulty motor or
Check motor and
leadscrew drive
leadscrew drive
assembly.
assembly.
5
Power supply.
Faulty power supply.
Check the power supply.
Code
Fault
Recommended action
78
Interference or internal circuit
See item 2 below
79
Interference or internal circuit
See item 2 below
80
Interference or internal circuit
See item 2 below
81
Interference or internal circuit
See item 2 below
83
Interference or internal circuit
See item 2 below
85
Power supply fault
See item 5 below
86
Motor fault
See item 4 below
87
Interference or internal circuit
See item 2 below
88
Interference or internal circuit
See item 2 below
90
Interference or internal circuit
See item 2 below
91
Keyboard fault
See item 3 below
92
Interference or internal circuit
See item 2 below
Table 5.2 Fault cause and action
Summary of Contents for Graseby 3200
Page 15: ...CHAPTER 1 INTRODUCTION FEATURES and SPECIFICATION 3200 In line Pressure Syringe Pump ...
Page 22: ...CHAPTER 2 CONFIGURATION DIAGNOSTICS and OCCLUSION THRUST 3200 In line Pressure Syringe Pump ...
Page 37: ...CHAPTER 3 FUNCTIONAL DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Page 43: ...CHAPTER 4 CIRCUIT DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Page 83: ...CHAPTER 6 FUNCTIONAL TESTS and MANUFACTURNG SETTINGS 3200 In line Pressure Syringe Pump ...
Page 91: ...CHAPTER 7 ILLUSTRATED PARTS LIST 3200 In line Pressure Syringe Pump ...
Page 105: ...CHAPTER 8 BRAUN PERFUSOR CONVERSION 3200 In line Pressure Syringe Pump ...
Page 108: ...CHAPTER 9 DC INPUT VERSION of 3200 3200 In line Pressure Syringe Pump ...
Page 111: ...APPENDIX FITTING a MODIFIED SIZE SENSOR FLAG 3200 In line Pressure Syringe Pump ...