Smiths Medical International Ltd.
3200 Service Manual
Issue 5 (August 2004)
2 — 4
Configuration mode
This parameter shows the versatility of the 3200 by allowing five different combina-
tions of the three modes of infusions to be pre-selected.
Choice:
ALL MODES AVAILABLE
CONTINUOUS & PRESET
INTERMITTENT ONLY
PRESET VOLUME ONLY
CONTINUOUS ONLY
Select pump
modes
5.
Preset volume
mode with
time
6.
This parameter allows either a preset volume with time (YES selected) or a preset
volume with rate (NO selected).
Choice:
YES or NO.
KVO rate
7.
This parameter allows either a KVO rate of 0.50 ml/h or 0.05 ml/h to be chosen.
Choice:
0.5 ml/h or 0.05 ml/h.
When
YES
is chosen this parameter allows any one of the six mass units to be
selected [see parameters (9) to (14)].
When
NO
is chosen this parameter allows only ml/h to be selected.
Choice:
YES or NO
Allow mass
units option
8.
When
YES
is chosen from parameter (8), parameters (9) to (14) allow any combination
of mass units from the following list:
mg/kg/h.
µg/kg/h.
mg/kg/min.
µg/kg/min.
mg/h.
µg/h.
Choice:
YES or NO for each parameter.
9.
10.
11.
12.
13.
14.
Infusion units
This parameter allows the infusion rate, when set to mass units, i.e.
YES
, to also be
displayed in ml/h when infusing.
Choice:
YES/NO
Show rate in
15.
ml/h while
infusing
(contd.)
Summary of Contents for Graseby 3200
Page 15: ...CHAPTER 1 INTRODUCTION FEATURES and SPECIFICATION 3200 In line Pressure Syringe Pump ...
Page 22: ...CHAPTER 2 CONFIGURATION DIAGNOSTICS and OCCLUSION THRUST 3200 In line Pressure Syringe Pump ...
Page 37: ...CHAPTER 3 FUNCTIONAL DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Page 43: ...CHAPTER 4 CIRCUIT DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Page 83: ...CHAPTER 6 FUNCTIONAL TESTS and MANUFACTURNG SETTINGS 3200 In line Pressure Syringe Pump ...
Page 91: ...CHAPTER 7 ILLUSTRATED PARTS LIST 3200 In line Pressure Syringe Pump ...
Page 105: ...CHAPTER 8 BRAUN PERFUSOR CONVERSION 3200 In line Pressure Syringe Pump ...
Page 108: ...CHAPTER 9 DC INPUT VERSION of 3200 3200 In line Pressure Syringe Pump ...
Page 111: ...APPENDIX FITTING a MODIFIED SIZE SENSOR FLAG 3200 In line Pressure Syringe Pump ...