36
ADVIA 1800 Operator’s Guide
1
Reference Electrode
6
Dilution Probe (DPP)
11
Buffer Pump (BP)
2
Na Electrode
7
Sample
12
Temperature Stabilizer
3
K Electrode
8
BPEV
13
DPEV
4
Cl Electrode
9
ISE Degassing Unit
14
DP
5
Mixer
10
Buffer Solution
15
Waste Block
Figure 1-27. ISE components
To ensure data accuracy, undiluted sample is aspirated from the sample tray and
electrolyte analysis is always performed before photometric sampling.
Universal rack handler – functional description (optional)
The universal rack handler is a separate sample delivery mechanism designed to allow
operators to continuously load samples. The operator can load five sample containers into
each rack and up to fifteen racks (75 samples) on each rack carrier to improve workflow
efficiency. The sample-dilution probe (DPP) aspirates the sample on the rack handler just
as it does from the sample tray, from an aspiration position specifically designed for the
universal rack handler or another laboratory automation system (LAS).
Use the universal rack handler for routine patient sample analysis only.
Process calibrators, controls, special diluents, reruns, and STAT samples with the onboard
sample tray (STT).
Summary of Contents for ADVIA 1800
Page 102: ...102 ADVIA 1800 Operator s Guide ...
Page 108: ...108 ADVIA 1800 Operator s Guide ...
Page 182: ...182 ADVIA 1800 Operator s Guide ...
Page 188: ...188 ADVIA 1800 Operator s Guide ...
Page 194: ...194 ADVIA 1800 Operator s Guide ...
Page 198: ...198 ADVIA 1800 Operator s Guide ...
Page 204: ...204 ADVIA 1800 Operator s Guide ...
Page 212: ...212 ADVIA 1800 Operator s Guide ...