Maintenance
147
4.
Immerse the cuvette segments in 5% Probe Wash 3 solution.
a.
Ensure no air bubbles are in the cuvettes.
b.
Allow the cuvettes to soak for at least 10 hours.
5.
Wash the cuvettes under running water, then rinse them in deionized water.
6.
Drain the water from the cuvettes.
7.
Install the cuvette segments on the dilution tray (DTT) and fasten the thumbscrews
by hand.
8.
Return the analyzer to the READY mode.
9.
At the Operation Panel, select
Initialize
.
NOTE:
Verify that the Operating mode field displays READY before performing any
further actions.
10.
At the Operation Panel, select
WASH
.
11.
Perform the daily shutdown wash routine, then verify the operation.
Cleaning and replenishing the cuvette conditioner bottle
Materials required:
• Deionized water
• Cuvette conditioner
Time: 10 minutes
Analyzer mode: STANDBY
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
NOTE:
The cuvette conditioner bottle may be cleaned when it is refilled, but must be
cleaned at least once every 2 months.
WARNING
Before handling any reagents, read the warnings on page 196
1
Cuvette Conditioner Bottle
2
Level Sensor Connector
3
Filter Cap
Figure 5-21. Cuvette conditioner bottle
Summary of Contents for ADVIA 1800
Page 102: ...102 ADVIA 1800 Operator s Guide ...
Page 108: ...108 ADVIA 1800 Operator s Guide ...
Page 182: ...182 ADVIA 1800 Operator s Guide ...
Page 188: ...188 ADVIA 1800 Operator s Guide ...
Page 194: ...194 ADVIA 1800 Operator s Guide ...
Page 198: ...198 ADVIA 1800 Operator s Guide ...
Page 204: ...204 ADVIA 1800 Operator s Guide ...
Page 212: ...212 ADVIA 1800 Operator s Guide ...