150
ADVIA 1800 Operator’s Guide
3.
Loosen the lead wire connectors (
1
and
2
), then remove the wires.
4.
Unfasten the lamp screws (
3
) on the plate (
4
) and remove the halogen lamp from the
housing.
CAUTION
Be careful not to drop the screws.
5.
When installing the new lamp, align the hole (
1
) to the locating pin (
2
).
Figure 5-23. Locating pin
CAUTION
Do not touch the glass portion of the lamp. Touching the lamp could damage it. If
the lamp is dirty, clean it using lint-free toweling moistened with ethanol.
6.
Install the lamp screws.
7.
Install the lead wires and fasten the knobs.
8.
Replace the access panel.
9.
Return the analyzer to the
READY
mode, if necessary.
10.
Wait 40 minutes for the lamp to stabilize.
NOTE:
You can perform the daily shutdown 2 while you are waiting.
11.
Check the lamp energy (see the
Weekly Maintenance
section).
12.
Register the data and verify that attenuations register at 100%, voltages are between
5–9, and scatter is no greater than ±40.
13.
Perform the cell blank measurement test.
NOTE:
Siemens recommends the assays on the system be calibrated after the lamp is
replaced.
14.
Run QC controls to verify that all assays are within the laboratory’s established
control ranges.
Summary of Contents for ADVIA 1800
Page 102: ...102 ADVIA 1800 Operator s Guide ...
Page 108: ...108 ADVIA 1800 Operator s Guide ...
Page 182: ...182 ADVIA 1800 Operator s Guide ...
Page 188: ...188 ADVIA 1800 Operator s Guide ...
Page 194: ...194 ADVIA 1800 Operator s Guide ...
Page 198: ...198 ADVIA 1800 Operator s Guide ...
Page 204: ...204 ADVIA 1800 Operator s Guide ...
Page 212: ...212 ADVIA 1800 Operator s Guide ...