
166
ADVIA 1800 Operator’s Guide
8.
Snap the joint holder (
3
) back into the post.
9.
Replace the splash guards and secure them back in place with the thumb screws.
10.
Manually lift and rotate the probe over the probe wash cup but not within the wash
port.
11.
Replace the DPP shield and secure it in place.
Aligning the DPP probe
1.
Return the analyzer to the READY mode.
2.
At the Menu Panel, select
Maint
, then select
User Maintenance
.
3.
In the Probe posi.adjust area, select Position probes for routine cleaning to move all
probes (DPP, SPP, RPP1, and RPP2) over cuvettes.
DPP should be over cuvette 1 on the DTT.
4.
Ensure that the probe is perpendicular to the arm and centered over the cuvette.
If it is not, call your local technical support provider or distributor.
5.
At the Operation Panel, select
Prime
, then
PRIME 2
, then
Execute
to ensure proper
water flow through the probe.
6.
Verify that water is not leaking from the joint connector (
2
).
Replenishing the RRV (reaction) bath oil bottle
Materials required:
• RRV (reaction) bath oil
(REF 09323099, PN B01-4180-01)
Time: 10 minutes
Analyzer mode: STANDBY
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
1
RRV (Reaction) Bath Oil
Bottl
e
2
Level Sensor Connector
3
Filter Cap
Figure 5-33. RRV bath oil bottle
Summary of Contents for ADVIA 1800
Page 102: ...102 ADVIA 1800 Operator s Guide ...
Page 108: ...108 ADVIA 1800 Operator s Guide ...
Page 182: ...182 ADVIA 1800 Operator s Guide ...
Page 188: ...188 ADVIA 1800 Operator s Guide ...
Page 194: ...194 ADVIA 1800 Operator s Guide ...
Page 198: ...198 ADVIA 1800 Operator s Guide ...
Page 204: ...204 ADVIA 1800 Operator s Guide ...
Page 212: ...212 ADVIA 1800 Operator s Guide ...