Maintenance
167
CAUTION
Do not attempt to clean the RRV bath oil bottle (
1
) with water; RRV bath oil
and water do not mix.
1.
Set the analyzer to Standby.
2.
Unscrew the filter cap (
3
) at the front of the RRV bath oil bottle (
1
), then pull up the
tube with the filter.
3.
Disconnect the RRV bath oil bottle level sensor connector (
2
), then turn the
connector counter-clockwise and pull it out.
CAUTION
Make a note of the bottle position on the shelf to avoid mixing up the fluid
bottles.
4.
Remove the RRV bath oil bottle (
1
).
CAUTION
Ensure that level-sensor connector (
2
) does not get wet to avoid damaging it.
5.
Refill the bottle with RRV (reaction) bath oil.
6.
Replace the bottle on the shelf in the cabinet.
7.
Connect the RRV bath oil bottle level sensor connector (
2
) by pushing the connector
in and turning it clockwise.
8.
Insert the filter and tube, then fasten the cap.
NOTE:
Make sure that the filter holder is located at the bottom of the bottle.
9.
Return the analyzer to the READY mode.