System Overview
33
During each half-turn (3-second cycle), the nozzles operate as follows:
Nozzle
Probe
Description
1
A
Aspirates reaction liquid.
B1
Dispenses wash water.
H1
Absorbs overflow liquid (in abnormal conditions).
2
C1
Aspirates wash water.
D
Dispenses detergent.
H2
Absorbs overflow liquid (in abnormal conditions).
3
E
Aspirates detergent.
B2
Dispenses wash water.
H3
Absorbs overflow liquid (in abnormal conditions).
4
C2
Aspirates wash water.
F
Dispenses cell conditioner.
H4
Absorbs overflow liquid (in abnormal conditions).
5
G
Aspirates cell conditioner.
B3
Dispenses wash water.
H5
Absorbs overflow liquid (in abnormal conditions).
6
C3
Aspirates wash water.
7
I
Vacuums remaining liquid from cuvette.
Figure 1-24. Nozzle separation
If an assay is not running, you can operate the unit manually at the Manual Operation
window.
Summary of Contents for ADVIA 1800