210
ADVIA 1800 Operator’s Guide
Item
Description
Optional universal rack handler
dimensions
95.25(h) x 72.5(w) x 103.0(d) cm, (37.5 x 28.5 x 40.55 in.)
Weight
80.7 kg (178 lb)
Environment
Electrical requirements
•
The system, with or without the rack handler, conforms to
Installation/Overvoltage Category II.
•
Space around the system must be sufficient to allow for ventilation.
There must be a minimum of 19.7 inches (500 mm) behind the system,
32 inches (800 mm) on each side, and 42 inches (1000 mm) in front of
the system.
•
The system should not be exposed to direct sunlight.
•
Environment should be free of corrosive, flammable, or anesthetic gases
(the analyzer is not built to be explosion-proof); significant vibration;
and electrical disturbances, such as electromagnetic and electrostatic
induction.
•
The system should be placed on a level floor (1/200 gradient or less)
that is capable of supporting a load of 560 kg (1234 lb).
•
The average acoustic noise output from the analyzer is <70 dba with the
top cover open.
•
A 3-kVA power source, single-phase, 2-pole, 3-wire configuration with
Class III grounding
•
The following input voltages can be tapped: 200, 220, 230, or 240 VAC
at 50/60 Hz.
•
Main supply voltage fluctuations not to exceed ± 10 percent of the
nominal voltage. Maximum current draw at in-rush is 15 amps at 200
VAC or 13 amps at 240 VAC.
•
For the optional universal rack handler, the requirements are 110 Vac
50/60 Hz, 0.6 A.
•
Facility Switch Box: Should contain a circuit breaker and knife switch
with one of the following ratings:
200 volts 15 amps, 220 volts 14 amps, 230 volts 13 amps, and 240 volts
13 amps
•
Switch box should be located less than 5 m (15 feet) from the system
and easily accessible.
NOTE:
Mark near the circuit breaker and knife switch that the switch
box is a disconnecting device for the equipment.
•
Facility Supply Wiring: UL listed cord for external wiring in US and
Canada. (Must be CSA certified in Canada.)
Rated 300 volts or more, and 70
°
C or more
Size AWG # 14 or more (OD 13 to 18 mm), 3-wire configuration that
has a protective ground wire with covering material colored with
green and yellow stripe.
Circuit breaker - 15 or 20 amps.
Cooling/ventilation requirements
•
Ventilation sufficient to ma18
°
C to +30
°
C (+64
°
to +86
°
F)
operating temperature
•
The maximum temperature change the system can accommodate is
2
°
C/hour.
Summary of Contents for ADVIA 1800
Page 102: ...102 ADVIA 1800 Operator s Guide ...
Page 108: ...108 ADVIA 1800 Operator s Guide ...
Page 182: ...182 ADVIA 1800 Operator s Guide ...
Page 188: ...188 ADVIA 1800 Operator s Guide ...
Page 194: ...194 ADVIA 1800 Operator s Guide ...
Page 198: ...198 ADVIA 1800 Operator s Guide ...
Page 204: ...204 ADVIA 1800 Operator s Guide ...
Page 212: ...212 ADVIA 1800 Operator s Guide ...