220
ADVIA 1800 Operator’s Guide
error
temperature is outside the
pre-defined range
G
Crash
A probe crash is detected
Y
Y
N
h, l
Normal value
(upper,
lower)
Result exceeds expected
range
Y
Y
Y
H, L
Abnormal
value
(upper,
lower)
Result exceeds expected
range
Y
Y
Y
i
Ratio results
Ratio component test is
flagged
Y
Y
j
Exceeded the
calibration
high range
The calibration high range is
exceeded.
Y
Y
K
Maximum
absorbance
limit
exceeded
The maximum absorbance
limit is exceeded.
NOTE
This flag is linked to
the Max Limit field at the
Analytical Parameters
Chemistry window.
Y
Y
k
Exceeded the
calibration
low range
The calibration low range is
exceeded.
Y
Y
M
Mix error
A mixer motion error is
detected
Y
Y
N
n
Abnormal
number of
effective
points
Insufficient valid data points
Y
Y
N
N
Cell blank
Abnormal cell blank
N
Y
N
P
Prozone
Failed prozone check
Y
Y
N
Q
Liquid level
sensor error
A liquid level sensor error
occurs
Y
Y
__
s
Photometric test insufficient
sample
Y
Y
N
Summary of Contents for ADVIA 1800
Page 102: ...102 ADVIA 1800 Operator s Guide ...
Page 108: ...108 ADVIA 1800 Operator s Guide ...
Page 182: ...182 ADVIA 1800 Operator s Guide ...
Page 188: ...188 ADVIA 1800 Operator s Guide ...
Page 194: ...194 ADVIA 1800 Operator s Guide ...
Page 198: ...198 ADVIA 1800 Operator s Guide ...
Page 204: ...204 ADVIA 1800 Operator s Guide ...
Page 212: ...212 ADVIA 1800 Operator s Guide ...