126
ADVIA 1800 Operator’s Guide
1.
Set the analyzer to Standby.
2.
Inspect the SP and DIP vertical pumps:
NOTE:
Liquid leaking from the seal on the sample aspiration pump (sp) or the dilution
aspiration pump (dip) flows to the drive lever unit. if the drive lever unit is wet, the
pump seal must be replaced. contact your local technical support provider or
distributor.
a.
Closely
inspect the following for
moisture:
•
Tubing connections (
1
)
•
Plastic cylinders (
2
)
•
Drip tray under the pumps
b.
If moisture is detected, ensure each tubing
connection is snug (hand-tightened).
c.
If leaks persist, call your local technical
support provider or distributor
Figure 5-10. Pump plastic cylinder
3.
Inspect the other pumps (not SP and DIP) for leaks:
The other pumps (not SP or DIP) are:
•
Dilution pump (DCP)
•
Dilution discharge pump (DOP)
•
Sample and reagent wash pump (SRWP)
•
Reagent dispensing pumps (RP1 and RP2)
a.
Closely
inspect the following for moisture:
•
Upper portion
(cylinders, L-ring
holders, tubes and fittings) of the
pumps (
1
).
•
Drip tray under the pumps
Figure 5-11. Typical Pump
b.
If moisture is detected, ensure each tubing connection is snug (hand-tightened).
c.
If leaks persist, call your local technical support provider or distributor.
Summary of Contents for ADVIA 1800
Page 102: ...102 ADVIA 1800 Operator s Guide ...
Page 108: ...108 ADVIA 1800 Operator s Guide ...
Page 182: ...182 ADVIA 1800 Operator s Guide ...
Page 188: ...188 ADVIA 1800 Operator s Guide ...
Page 194: ...194 ADVIA 1800 Operator s Guide ...
Page 198: ...198 ADVIA 1800 Operator s Guide ...
Page 204: ...204 ADVIA 1800 Operator s Guide ...
Page 212: ...212 ADVIA 1800 Operator s Guide ...