174
ADVIA 1800 Operator’s Guide
Priming and initializing the ISE module
1.
At the ISE Operation window, in the Bufferprime area, enter
15
in the Times field.
2.
To prime the line with buffer, select
Execute
, then select
Yes
.
3.
When the priming is finished, verify that the electrodes are not leaking.
4.
At the ISE Operation window, select
Exit
, then select
Yes
.
5.
Run 10 pooled serum samples, or do an ISE CV check.
6.
Perform calibration and run controls.
Replacing ISE electrodes
Materials required:
Electrodes
•
Cl (REF 07097504, PN 073-0049-01)
•
K (REF 06135445, PN 073-0050-01)
•
Na (REF 03092699, PN 073-0051-01)
•
Reference (REF 00311764, PN 073-0653-01
•
O-rings, 3 (REF 09955206, PN 073-0071-01)
•
Philips screwdriver
Time: 5 minutes
Analyzer mode: Manual operation
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
Replace the Na, K, and the Cl electrodes if the slope is incorrect or calibration
continuously fails.
The acceptable ISE slope is between 45.0 and
63.0. Slopes outside of this range are flagged
as shown in the table. A flagged slope fails
the calibration. The slope limits are defined at
the ISE Parameter Settings window.
Mark
ISE Slope Range
H
> 65.0
h
63.1 to 65.0
l
38.0 - 44.9
L
< 38.0
Replace the reference electrode when the reference electrode value is <500.
Checking the reference electrode value
1.
At the Menu Panel, select
Maint
., then select
ISE Monitor
.
2.
At the ISE Monitor window, at the bottom of the Calib.monitor: Serum area, check
the value of the Ref. electrode field.
3.
If the Ref. electrode value is
less than 500.0
, replace the reference electrode.
Removing electrodes
1.
At the Menu Panel, select
Maint.
, then select
ISE Operation
.
2.
In the Period. wash area, select
OFF
, then select
Set
.
3.
Using a Phillips screwdriver, remove the screws (
1
, see Figure 5-2) that secure the
DPP shield to the analyzer panel.
Summary of Contents for ADVIA 1800
Page 102: ...102 ADVIA 1800 Operator s Guide ...
Page 108: ...108 ADVIA 1800 Operator s Guide ...
Page 182: ...182 ADVIA 1800 Operator s Guide ...
Page 188: ...188 ADVIA 1800 Operator s Guide ...
Page 194: ...194 ADVIA 1800 Operator s Guide ...
Page 198: ...198 ADVIA 1800 Operator s Guide ...
Page 204: ...204 ADVIA 1800 Operator s Guide ...
Page 212: ...212 ADVIA 1800 Operator s Guide ...