Maintenance
173
Figure 5-37. ISE electrode plate thumbscrew
7.
Remove the electrodes and replace it with the dummy electrode.
8.
At the ISE Operation window, in the Bufferprime area, type
50
in the Times field.
9.
Select
Execute
10.
When prompted, select
Yes
to execute buffer prime.
Washing the lines
1.
Remove the buffer bottle with the deionized water and replace it with a bottle filled
with a solution of 475 mL of deionized water and 25 mL probe wash 3 solution.
2.
At the ISE Operation window, in the Bufferprime area, enter
50
in the Times field.
3.
Select
Execute
.
4.
When prompted, select
Yes
to execute the buffer prime.
Rinsing the lines
1.
Replace the probe wash 3 solution bottle with a bottle of deionized water.
2.
In the Bufferprime area, enter
50
in the Times field, then select
Execute
.
3.
Remove the dummy electrode.
4.
Reinstall the Na, K, and Cl electrodes.
5.
In the Initialize area, select
Execute
.
6.
Before reinstalling the buffer-solution bottle, thoroughly rinse the buffer bottle cap,
float switch, and tube with deionized water and dry completely.
7.
Install the buffer bottle or replace it if the volume is low.
Summary of Contents for ADVIA 1800
Page 102: ...102 ADVIA 1800 Operator s Guide ...
Page 108: ...108 ADVIA 1800 Operator s Guide ...
Page 182: ...182 ADVIA 1800 Operator s Guide ...
Page 188: ...188 ADVIA 1800 Operator s Guide ...
Page 194: ...194 ADVIA 1800 Operator s Guide ...
Page 198: ...198 ADVIA 1800 Operator s Guide ...
Page 204: ...204 ADVIA 1800 Operator s Guide ...
Page 212: ...212 ADVIA 1800 Operator s Guide ...