148
ADVIA 1800 Operator’s Guide
1.
Set the analyzer to Standby.
2.
Open the filter cap at the front of the cuvette conditioner bottle (
3
), and pull up the
tube with the filter.
3.
Disconnect the cuvette conditioner bottle level sensor connector (
2
).
4.
Turn the connector counter-clockwise and pull it out.
CAUTION
Make a note of the bottle position on the shelf to avoid mixing up the fluid
bottles.
5.
Remove the bottle.
6.
Empty any remaining contents of the bottle.
7.
Rinse the bottle with deionized water and drain well.
CAUTION
Ensure that the level-sensor connector (
2
) does not get wet to avoid damaging it.
8.
Refill the bottle with cuvette conditioner.
9.
Return the bottle to the same position on the shelf in the cabinet.
10.
Connect the cuvette-conditioner bottle level sensor connector, then push the
connector in and turn clockwise.
11.
Insert the filter and hose, then fasten the cap.
NOTE:
Make sure that the filter holder is located at the bottom of the bottle.
12.
Prime the fluid lines:
NOTE:
If you are cleaning other bottles, wait to perform this step for all fluid lines.
a.
Return the analyzer to the READY mode.
b.
At the Operation Panel, select
Prime
.
c.
At the Prime Settings dialog box, select
Prime 2
and type
10
for the Number of
times in all fields.
d.
Select
Execute
.
Summary of Contents for ADVIA 1800
Page 102: ...102 ADVIA 1800 Operator s Guide ...
Page 108: ...108 ADVIA 1800 Operator s Guide ...
Page 182: ...182 ADVIA 1800 Operator s Guide ...
Page 188: ...188 ADVIA 1800 Operator s Guide ...
Page 194: ...194 ADVIA 1800 Operator s Guide ...
Page 198: ...198 ADVIA 1800 Operator s Guide ...
Page 204: ...204 ADVIA 1800 Operator s Guide ...
Page 212: ...212 ADVIA 1800 Operator s Guide ...