132
ADVIA 1800 Operator’s Guide
c.
Type
49
in the CTT cup position 1st time field and
50
in the CTT cup position
2nd time field.
d.
Type
55
in the RTT1 and RTT2 cup positions 1st time fields and
56
in the RTT1
and RTT2 cup positions 2nd time fields.
7.
Select
Execute
.
8.
When the wash is complete, remove the 5% probe wash 3 solution from position 55
on RTT1 and RTT2, and replace them with 10% cuvette wash solution.
9.
After the wash, check the lamp energy and run the cell blank measurement test.
NOTE:
Perform the lamp energy check and cuvette blank measurement only once a
week, even if the weekly wash is run daily.
Checking and replenishing the lamp coolant
Materials required:
• Lamp coolant additive (REF
04533710, PN B01-4496-51)
• Deionized water
Time: 5 minutes
Analyzer mode: READY
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
The lamp is cooled by circulating liquid coolant. As the volume of coolant decreases, the
heat of the lamp increases.
NOTE:
Check the lamp coolant level daily and whenever the system generates a lamp
coolant warning and turns off the lamp.
1.
Lift off the lamp access cover (
1
)
to gain access to the lamp coolant reservoir.
Figure 5-13. Lamp access cover
2.
Check the fluid level in the reservoir.
If the level is between the 5- and 9-cm marks, the label is green on the reservoir, proceed
to step 4.
Summary of Contents for ADVIA 1800
Page 102: ...102 ADVIA 1800 Operator s Guide ...
Page 108: ...108 ADVIA 1800 Operator s Guide ...
Page 182: ...182 ADVIA 1800 Operator s Guide ...
Page 188: ...188 ADVIA 1800 Operator s Guide ...
Page 194: ...194 ADVIA 1800 Operator s Guide ...
Page 198: ...198 ADVIA 1800 Operator s Guide ...
Page 204: ...204 ADVIA 1800 Operator s Guide ...
Page 212: ...212 ADVIA 1800 Operator s Guide ...