LF(R) Series Operation Manual
3 Various Measuring Methods
- 35 -
10
Place the sample in the cage.
Place the sample in the cage and immerse the sample
fully in the water.
When the weight readout is stabilized, press the [Set]
key.
The measured specific gravity value is displayed.
Keep the cage away from the liquid container.
Pressing the [Function] key toggles the display
between the specific gravity and the volume.
Press the [Set] key to return to the weight display.
3-5-4 When using a liquid other than water
1
Go to the function setting mode.
(Refer to Section 2-6 “Basics of Function
Settings.”)
Press and hold the [Function] key until “Func” is
displayed.
2
Select gravimeter mode.
Press the [Zero/Tare] key several times to select “1.SEt
5.”
3
Go to the medium selection mode.
Press the [Function] key.
“11.MEd.” is displayed.
4
Select the “Not water” setting.
Press the [Zero/Tare] key to select “1” (Not water), and
press the [Set] key.
Now the balance can measure the specific
gravity using any liquid other than water.
Measure the specific gravity with the steps
described in “Measurement” (
page
エラー! ブッ
クマークが定義されていません。
). In step 5, set
the density of the liquid you are using, instead of
the water temperature. The liquid density can be
set to between 0.0001 and 9.999 g/cm
3
.
Summary of Contents for LF124R
Page 130: ...7 Troubleshooting LF R Series Operation Manual 120 Memo ...
Page 131: ...LF R Series Operation Manual Appendixes 121 Appendixes Appendix 1 Function Setting List ...
Page 132: ...Appendixes LF R Series Operation Manual 122 ...
Page 133: ...LF R Series Operation Manual Appendixes 123 ...
Page 134: ...Appendixes LF R Series Operation Manual 124 ...
Page 135: ...LF R Series Operation Manual Appendixes 125 ...
Page 136: ...Appendixes LF R Series Operation Manual 126 ...
Page 137: ...LF R Series Operation Manual Appendixes 127 ...
Page 138: ...Appendixes LF R Series Operation Manual 128 ...
Page 139: ...LF R Series Operation Manual Appendixes 129 ...
Page 144: ...Appendixes LF R Series Operation Manual 134 α check Display test α check Key test ...
Page 145: ...LF R Series Operation Manual Appendixes 135 α check Motor test Span adjustment history ...
Page 149: ...LF R Series Operation Manual Appendixes 139 Appendix 4 4 Unit Conversion Table ...
Page 150: ...Appendixes LF R Series Operation Manual 140 Appendix 4 5 Outline Drawings ...
Page 153: ......
Page 154: ......