6 Input/Output to/from External Devices
LF(R) Series Operation Manual
- 104 -
6-5 Input Commands
Use these commands to control the balance from an external device.
(1) Tare range setting command, (2) Output control setting command, (3) Measurement
mode setting command, (4) Date output request and time output request command, (5)
Interval time setting command, (6) Span adjustment and span test command, and (7)
Limit value setting command
6-5-1 Procedure for transmission
(1) An input command is sent from an external device to the balance.
The full-duplex transmission system allows you to send input commands at any time
regardless of the data transmit timing of the balance.
(2) Upon successful completion of an input command, the balance will send either a
normal completion response or the result data requested by the command to the
external device.
If the operation has not resulted in successful completion, or if the command is invalid
(an error), the balance will transmit an error response.
When the balance is in normal display mode, it usually sends a response to a
command within one second of receiving the command. For the tare range, span
adjustment or span test commands, a response is sent after the commands are
completely processed.
If the balance receives a tare range setting command when the function setting for
tare range (function setting H.tr) is set to “wait for the balance to stabilize for
zero-point adjustment (waiting for the balance to stabilize),” or if the balance receives
an input command that takes a long time, the balance sends a response after the
command is completely processed.
If the balance receives a command when you are setting a function, when the balance
is under span adjustment, or the balance is busy for other reasons, the command is
executed after the current operation has been completed.
After you have sent an input command, do not send another command to the
balance until the external device receives a response from the balance.
6-5-2 Input command examples
Sent command
Description
T(SP)(CR)(LF)
Set the tare range (adjust the zero point).
O1 (CR)(LF)
Set to continuous output.
O8 (CR)(LF)
Output data (once immediately).
Summary of Contents for LF124R
Page 130: ...7 Troubleshooting LF R Series Operation Manual 120 Memo ...
Page 131: ...LF R Series Operation Manual Appendixes 121 Appendixes Appendix 1 Function Setting List ...
Page 132: ...Appendixes LF R Series Operation Manual 122 ...
Page 133: ...LF R Series Operation Manual Appendixes 123 ...
Page 134: ...Appendixes LF R Series Operation Manual 124 ...
Page 135: ...LF R Series Operation Manual Appendixes 125 ...
Page 136: ...Appendixes LF R Series Operation Manual 126 ...
Page 137: ...LF R Series Operation Manual Appendixes 127 ...
Page 138: ...Appendixes LF R Series Operation Manual 128 ...
Page 139: ...LF R Series Operation Manual Appendixes 129 ...
Page 144: ...Appendixes LF R Series Operation Manual 134 α check Display test α check Key test ...
Page 145: ...LF R Series Operation Manual Appendixes 135 α check Motor test Span adjustment history ...
Page 149: ...LF R Series Operation Manual Appendixes 139 Appendix 4 4 Unit Conversion Table ...
Page 150: ...Appendixes LF R Series Operation Manual 140 Appendix 4 5 Outline Drawings ...
Page 153: ......
Page 154: ......