LF(R) Series Operation Manual
3 Various Measuring Methods
- 27 -
3-3 Measuring Percentage
The weight of a sample is shown as a percentage of the reference sample weight.
A reference sample weight can be set by weighing an actual sample (setting a reference
weight by weighing an actual sample) or entering a value (setting a reference weight by
entering a value).
3-3-1
Setting a reference weight by weighing an actual sample
1
Go to the function setting mode.
(Refer to Section 2-6 “Basics of Function
Settings.”)
Press and hold the [Function] key until “Func” is
displayed.
2
Select percentage weighing mode.
Press the [Zero/Tare] key several times to select
“1. SEt 3.”
3
Save the mode.
Press the [Set] key.
The percentage weighing mode is set and “%” is
displayed.
4
Start measuring the reference sample weight.
<Press and hold>
Press and hold the [Function] key. After “P. SEt” is
displayed, release the key.
The previously saved reference sample weight
blinks.
5
Save the reference sample weight.
Put the reference sample on the balance and then press
the [Function] key.
The reference weight is saved.
6
Weigh a sample.
The display indicates the percentage of the sample to the
reference sample weight.
Pressing the [Function] key toggles the display
between percentage weight and total weight.
Summary of Contents for LF124R
Page 130: ...7 Troubleshooting LF R Series Operation Manual 120 Memo ...
Page 131: ...LF R Series Operation Manual Appendixes 121 Appendixes Appendix 1 Function Setting List ...
Page 132: ...Appendixes LF R Series Operation Manual 122 ...
Page 133: ...LF R Series Operation Manual Appendixes 123 ...
Page 134: ...Appendixes LF R Series Operation Manual 124 ...
Page 135: ...LF R Series Operation Manual Appendixes 125 ...
Page 136: ...Appendixes LF R Series Operation Manual 126 ...
Page 137: ...LF R Series Operation Manual Appendixes 127 ...
Page 138: ...Appendixes LF R Series Operation Manual 128 ...
Page 139: ...LF R Series Operation Manual Appendixes 129 ...
Page 144: ...Appendixes LF R Series Operation Manual 134 α check Display test α check Key test ...
Page 145: ...LF R Series Operation Manual Appendixes 135 α check Motor test Span adjustment history ...
Page 149: ...LF R Series Operation Manual Appendixes 139 Appendix 4 4 Unit Conversion Table ...
Page 150: ...Appendixes LF R Series Operation Manual 140 Appendix 4 5 Outline Drawings ...
Page 153: ......
Page 154: ......