3 Various Measuring Methods
LF(R) Series Operation Manual
- 34 -
6
Save the water temperature.
Press the [Set] key.
7
Hang the cage only and set the weight readout to
zero.
Hang only the cage from the underfloor weighing hook.
Press the [Zero/Tare] key.
The readout changes to zero.
8
Weigh a sample.
Place a sample in the cage.
When the weight readout is stabilized, press the [Set]
key.
This saves the weight of the sample in the air.
Once the weight is saved, “ ” is shown in the
lower left corner of the display.
You can also weigh the sample on the weighing
pan.
9
Set the liquid container in place.
Place the container with water under the balance, and
put only the cage into the water. (Do not put the sample
in the cage.)
Press the [Zero/Tare] key to change the readout to zero.
This eliminates the discrepancy caused by the
cage.
Summary of Contents for LF124R
Page 130: ...7 Troubleshooting LF R Series Operation Manual 120 Memo ...
Page 131: ...LF R Series Operation Manual Appendixes 121 Appendixes Appendix 1 Function Setting List ...
Page 132: ...Appendixes LF R Series Operation Manual 122 ...
Page 133: ...LF R Series Operation Manual Appendixes 123 ...
Page 134: ...Appendixes LF R Series Operation Manual 124 ...
Page 135: ...LF R Series Operation Manual Appendixes 125 ...
Page 136: ...Appendixes LF R Series Operation Manual 126 ...
Page 137: ...LF R Series Operation Manual Appendixes 127 ...
Page 138: ...Appendixes LF R Series Operation Manual 128 ...
Page 139: ...LF R Series Operation Manual Appendixes 129 ...
Page 144: ...Appendixes LF R Series Operation Manual 134 α check Display test α check Key test ...
Page 145: ...LF R Series Operation Manual Appendixes 135 α check Motor test Span adjustment history ...
Page 149: ...LF R Series Operation Manual Appendixes 139 Appendix 4 4 Unit Conversion Table ...
Page 150: ...Appendixes LF R Series Operation Manual 140 Appendix 4 5 Outline Drawings ...
Page 153: ......
Page 154: ......