87
Set the lens voltages with the Peak Monitor and confirm the output voltages at the
feedthrough contacts on the IS CTRL-A PCB.
The normal output voltage lies in the range: set voltage
±
5%.
LENS1, LENS2, LENS3, LENS4
If the output voltages do not lie within the set ranges, conduct the checks described in
2.
Checking the lens voltages
for MAIN CTRL-A (page 60).
3. Checking the filament currents
When a filament is on, the filament current can be determined from the following voltage
measurements. The relationship between the filament current and the voltage measured
with the probe on CP16[FIL] and the - probe on CP18[EV] is as follows:
Filament current (A) = 2 x (V
CP16
– V
CP18
) max about 3.5A
If (V
CP16
– V
CP18
) = 1.4V, the filament current is 2.8A.
A filament discontinuity is detected and an error occurs if (V
CP16
– V
CP18
)
≤
approx. 0.6V, that
is if the filament current
≤
approx. 1.2A.
The discontinuity in the filament can be easily confirmed by the following method.
Turn off the filament and measure the resistance between F1+ and F1-.
If resistance value
≤
1
Ω
, the filament is normal.
Be aware of the polarity when measuring the resistance; correct measurement is only
possible by applying a positive voltage to F1+ and a negative voltage to F1-. If the polarity
is unknown, turn off the instrument power, remove IS CTRL-A to eliminate the influence of its
circuits and measure the resistance directly at the feedthrough terminals.
4. Checking the TRAP current control circuit
Set the emission current, turn the filament on, and measure the voltages at CP12[EMC],
CP13[TRP], and CP14[+10A].
If V
CP14
≠
10V, the IS CTRL-A PCB is probably defective.
V
CP12
=set current(
µ
A)/100
If emission current = 60
µ
A, V
CP12
=0.6V.
Any abnormality suggests an incorrect D/A voltage from the CPU PCB.
V
CP12
=V
CP13
Any abnormality could be due to a filament discontinuity or incorrect installation.
5. Checking the emission voltage control circuit
This circuit generates the emission voltage (-10V to -200V).
Set the emission voltage, turn the filament on, and measure the voltages at CP15[EMV],
CP18[EV], and CP19[INV].
V
CP15
= 0.05 x set voltage (V)
If emission voltage = -70V, V
CP15
=-3.5V.
Summary of Contents for GCMS-QP2010 Series
Page 1: ...1 225 00612A Gas Chromatograph Mass Spectrometer GCMS QP2010 Service Manual...
Page 27: ...27...
Page 71: ...71 Wiring Diagrams 1 Wiring diagram Overall Wiring Diagrams...
Page 72: ...72 Wiring Diagrams 2 Wiring diagram MAIN POWER ASSY...
Page 73: ...73 Unit Operation Checks 3 Wiring diagram Details...
Page 76: ...76 MAIN POWER ASSY Block Diagram Parts inside the broken box are in the MAIN POWER ASSY...
Page 79: ...79 DC POWER A Block Diagram...
Page 84: ...84 MAIN CTRL A Block Diagram...
Page 85: ...85...
Page 90: ...90 IS CTRL A Block Diagram...
Page 94: ...94 HV PCB ASSY block diagram 1 UNNCI case of model EI CI 2 NCI case of model NCI...
Page 98: ...98 IG CTRL A PCB Block Diagram...
Page 109: ...109 DI CTRL A block diagram...
Page 150: ...126 Do not remove the 2 screws on the right side Remove the 2 screws on the left side...
Page 151: ...127 Remove cardboard Move the TW300 left by holding in both hands...
Page 155: ...131 Remove the cap of outlet Loosen the 3 fixing screws and remove the plate Remove the o ring...
Page 160: ...136 Remove the cardboard Completion of TMP mounting...
Page 161: ...137 12 Fix the flange to outlet Place the flange on the outlet Do not forget to place o ring...
Page 163: ...139 Slide fast the IS CTRL A board down and fix it 4 screws Connect the two connectors...
Page 164: ...140 14 Place the covers Place the front cover Place the front door...