83
4. Checking the Pirani gauge measurement control circuits
Check the operational amplifier power supply voltage.
CP9 +5V
Confirm that circuit output voltage is in the correct range.
CP8 0.47 to 2.84V (1 to 300Pa) (Voltage decreases as the pressure increases.)
CP8 0.07V max. Discontinuity in Pirani gauge.
Check the Pirani gauge resistance.
Disconnect the Pirani gauge cable from the MAIN CTRL-A PCB connector J5 and measure
the resistance between Pins 1 and 2. The normal resistance is 10 to 15
Ω
.
5. Checking the ion-gauge vacuum measurement circuit
If measurements seem incorrect, check the output from the ion-gauge collector current
amplifier circuit.
CP6 [IG1]
A degree of vacuum in the range 1.3x10
-3
to1.3x10
-1
Pa is determined from the voltage V
CP6
.
Degree of vacuum (Pa) = 3.33x10
-2
x V
CP6
When the degree of vacuum exceeds 1.3x10
-1
Pa (V
CP6
≥
4V), the filament is turned off for
protection.
CP7 [IG100]
A degree of vacuum below 1.3x10
-3
Pa is determined from the voltage V
CP7
.
Degree of vacuum (Pa) = 3.28x10
-4
x V
CP7
A discontinuity in the ion-gauge measurement bulb or connection cable is evaluated if V
CP7
is 0.1V or less, and the ion-gauge filament is turned off. Inspect the cable between the
measurement bulb and connector J4 on the MAIN CTRL-A PCB.
Check the ion-gauge control circuit (IG CTRL-A) if the ion-gauge measurement bulb filament
does not light.
Summary of Contents for GCMS-QP2010 Series
Page 1: ...1 225 00612A Gas Chromatograph Mass Spectrometer GCMS QP2010 Service Manual...
Page 27: ...27...
Page 71: ...71 Wiring Diagrams 1 Wiring diagram Overall Wiring Diagrams...
Page 72: ...72 Wiring Diagrams 2 Wiring diagram MAIN POWER ASSY...
Page 73: ...73 Unit Operation Checks 3 Wiring diagram Details...
Page 76: ...76 MAIN POWER ASSY Block Diagram Parts inside the broken box are in the MAIN POWER ASSY...
Page 79: ...79 DC POWER A Block Diagram...
Page 84: ...84 MAIN CTRL A Block Diagram...
Page 85: ...85...
Page 90: ...90 IS CTRL A Block Diagram...
Page 94: ...94 HV PCB ASSY block diagram 1 UNNCI case of model EI CI 2 NCI case of model NCI...
Page 98: ...98 IG CTRL A PCB Block Diagram...
Page 109: ...109 DI CTRL A block diagram...
Page 150: ...126 Do not remove the 2 screws on the right side Remove the 2 screws on the left side...
Page 151: ...127 Remove cardboard Move the TW300 left by holding in both hands...
Page 155: ...131 Remove the cap of outlet Loosen the 3 fixing screws and remove the plate Remove the o ring...
Page 160: ...136 Remove the cardboard Completion of TMP mounting...
Page 161: ...137 12 Fix the flange to outlet Place the flange on the outlet Do not forget to place o ring...
Page 163: ...139 Slide fast the IS CTRL A board down and fix it 4 screws Connect the two connectors...
Page 164: ...140 14 Place the covers Place the front cover Place the front door...