ZEBV-5020C
Customer to provide 11mm dia. connector for
direct connection of the exhaust hose.
Exhaust duct plant
Hose clamp 16mm
(standard accessory)
Vinyl tube 10mm ID x 4m
(standard accessory)
Hose clamp 16mm
(standard accessory)
Exhaust outlet
Rotary pump 1
Exhaust Duct Example
7. Other parts to be prepared
A Shimadzu service technician will adjust the instrument and measure inspection data
immediately after the instrument is installed. Please have the following required
items available at this time.
1. Capillary column DB-5ms, manufactured by J&W
ID 0.25mm, length 30m, film thickness 0.25
µ
m GLC Center number: 122-5532
2. Microsyringes for sample injection (e.g., 10
µ
L)
3. Standard sample for instrument evaluation
OFN 1pg (225-09558-02) for evaluation of EI, NCI(OPTION)
Benzophenone 100pg/ul (225-09558-02) for the evaluation of C I(OPTION)I
Cholestane 100ng/ul (225-09391-05) for the evaluation of DI I(OPTION)
4. Other service parts or consumables that the customer considers necessary.
See the
List of Service Parts and Consumables
for details.
Provided by customer
Summary of Contents for GCMS-QP2010 Series
Page 1: ...1 225 00612A Gas Chromatograph Mass Spectrometer GCMS QP2010 Service Manual...
Page 27: ...27...
Page 71: ...71 Wiring Diagrams 1 Wiring diagram Overall Wiring Diagrams...
Page 72: ...72 Wiring Diagrams 2 Wiring diagram MAIN POWER ASSY...
Page 73: ...73 Unit Operation Checks 3 Wiring diagram Details...
Page 76: ...76 MAIN POWER ASSY Block Diagram Parts inside the broken box are in the MAIN POWER ASSY...
Page 79: ...79 DC POWER A Block Diagram...
Page 84: ...84 MAIN CTRL A Block Diagram...
Page 85: ...85...
Page 90: ...90 IS CTRL A Block Diagram...
Page 94: ...94 HV PCB ASSY block diagram 1 UNNCI case of model EI CI 2 NCI case of model NCI...
Page 98: ...98 IG CTRL A PCB Block Diagram...
Page 109: ...109 DI CTRL A block diagram...
Page 150: ...126 Do not remove the 2 screws on the right side Remove the 2 screws on the left side...
Page 151: ...127 Remove cardboard Move the TW300 left by holding in both hands...
Page 155: ...131 Remove the cap of outlet Loosen the 3 fixing screws and remove the plate Remove the o ring...
Page 160: ...136 Remove the cardboard Completion of TMP mounting...
Page 161: ...137 12 Fix the flange to outlet Place the flange on the outlet Do not forget to place o ring...
Page 163: ...139 Slide fast the IS CTRL A board down and fix it 4 screws Connect the two connectors...
Page 164: ...140 14 Place the covers Place the front cover Place the front door...