64
Problems during tuning
8. No peaks obtained (peaks extremely small)
Check the tuning file.
□
Are the lens voltage values appropriate?
Revert the lens voltages to the default values if unclear whether the tuning values are
appropriate.
The default values are as follows:
L1 1.0V / L2 -20V / L3 -1.5V / L4 -40V / P-Rod -3.5V / M-Rod -3.5V / FIL-EV -70V /
FIL-EM 60V / CDD -10000V / EM -900 ~ -1200V
Check the introduction of PFTBA.
□
Is the PFTBA volume appropriate?
Approximately 5mm from the bottom is appropriate.
□
Is the sound of the PFTBA valve opening and closing audible?
If no sound of valve operation is audible
−−
> Check MAIN CTRL-A.
Check the lens voltage.
□
Is each lens voltage output correct?
If any trace of peaks is apparent, set each voltage individually to 0V and check for
changes to the peak height. If absolutely no peaks are visible, measure the
voltages at the checkpoints on MAIN CTRL-A.
Check the degree of vacuum.
□
Is the IG value within the normal range?
A value not exceeding 1 to 2E
-3
Pa is normal.
□
Appropriate GC column flow rate?
□
Large air components (m/z28, m/z32)?
See 6. Pirani gauge, IC value too high, IG does not light.
Summary of Contents for GCMS-QP2010 Series
Page 1: ...1 225 00612A Gas Chromatograph Mass Spectrometer GCMS QP2010 Service Manual...
Page 27: ...27...
Page 71: ...71 Wiring Diagrams 1 Wiring diagram Overall Wiring Diagrams...
Page 72: ...72 Wiring Diagrams 2 Wiring diagram MAIN POWER ASSY...
Page 73: ...73 Unit Operation Checks 3 Wiring diagram Details...
Page 76: ...76 MAIN POWER ASSY Block Diagram Parts inside the broken box are in the MAIN POWER ASSY...
Page 79: ...79 DC POWER A Block Diagram...
Page 84: ...84 MAIN CTRL A Block Diagram...
Page 85: ...85...
Page 90: ...90 IS CTRL A Block Diagram...
Page 94: ...94 HV PCB ASSY block diagram 1 UNNCI case of model EI CI 2 NCI case of model NCI...
Page 98: ...98 IG CTRL A PCB Block Diagram...
Page 109: ...109 DI CTRL A block diagram...
Page 150: ...126 Do not remove the 2 screws on the right side Remove the 2 screws on the left side...
Page 151: ...127 Remove cardboard Move the TW300 left by holding in both hands...
Page 155: ...131 Remove the cap of outlet Loosen the 3 fixing screws and remove the plate Remove the o ring...
Page 160: ...136 Remove the cardboard Completion of TMP mounting...
Page 161: ...137 12 Fix the flange to outlet Place the flange on the outlet Do not forget to place o ring...
Page 163: ...139 Slide fast the IS CTRL A board down and fix it 4 screws Connect the two connectors...
Page 164: ...140 14 Place the covers Place the front cover Place the front door...