ZEBV-5020C
Reagent gas
Please prepare the following thing as case, reagent gas with CI and also NCI.
Attaching, change 2 kinds of reagent gas simultaneously it can use it.
Please prepare it without fail because any gas of
isobutene or
methane are
necessary at the time of an
installation.
Gas
Purity Filling quantity
and container
capacity
Container with a
manufacturer
regurator
CH
4
metane
(017‑33021‑01)
99.95%
or more
Filling quantity
15MPa
Container capacity
3.4L
Takachiho
chemical
industry
TYPE
TV1024
Takachiho
chemical industry
TU1065(注)
Left screw
(040‑72004)
i―C
4
H
10
isobutane
(017‑33031‑03)
99.
9%
or more
Filling quantity
1.5kg
Container capacity
4.8L
Takachiho
chemical
industry
TYPE
4.8LP
Takachiho
chemical industry
TU‑1066(caution)
(225‑09552)
Adapter with
a/the container
(017‑33030)
NH
3
Ammonia
(017‑33010)
99.9%
or more
Filling quantity
3.5kg
Container capacity
10L
Sumitomo Seika
Chemicals
(Pay attention
container etc.
of note Nippon
Sanso because
there is in the
case of a right
screw)
Takachiho
chemical industry
TUS1063(caution)
Left screw
Stainless steel
production
(210‑14191)
There is two piece pipe of reagent gas as for CImodel and NCI model.
This type of reagent gas pipe can not connect to other reguretor of the abovetable.
Please consult it with a merchant, in the case that other types are used.
"Exit form: the
φ
8.5 hose joint hand is installed with a M12X1 bag nut and an exchange
possible type"
5. Gas
Helium (He) is employed as the carrier gas in the GCMS-QP2010.
The gas type, purity, and pressure are designated in the table below. Provide gas
and a gas pressure regulator of the types designated in the table, or their equivalents.
Gas Purity
Pressure and
capacity
Manufacturer
Pressure regulator
He
(helium)
99.995% min.
15MPa
(150kgf/cms) 47L
Iwatani & Co.
Ltd.
Koike Sanso Kogyo
Co., Ltd.
PPR-He
221-35999-03
* 10-digit numbers are Shimadzu catalog numbers.
300 to 980kPa He supply pressure is required to the GC. However, some supply
pressures have a limiting effect on the pressures that can be set at the GC. Normally
Summary of Contents for GCMS-QP2010 Series
Page 1: ...1 225 00612A Gas Chromatograph Mass Spectrometer GCMS QP2010 Service Manual...
Page 27: ...27...
Page 71: ...71 Wiring Diagrams 1 Wiring diagram Overall Wiring Diagrams...
Page 72: ...72 Wiring Diagrams 2 Wiring diagram MAIN POWER ASSY...
Page 73: ...73 Unit Operation Checks 3 Wiring diagram Details...
Page 76: ...76 MAIN POWER ASSY Block Diagram Parts inside the broken box are in the MAIN POWER ASSY...
Page 79: ...79 DC POWER A Block Diagram...
Page 84: ...84 MAIN CTRL A Block Diagram...
Page 85: ...85...
Page 90: ...90 IS CTRL A Block Diagram...
Page 94: ...94 HV PCB ASSY block diagram 1 UNNCI case of model EI CI 2 NCI case of model NCI...
Page 98: ...98 IG CTRL A PCB Block Diagram...
Page 109: ...109 DI CTRL A block diagram...
Page 150: ...126 Do not remove the 2 screws on the right side Remove the 2 screws on the left side...
Page 151: ...127 Remove cardboard Move the TW300 left by holding in both hands...
Page 155: ...131 Remove the cap of outlet Loosen the 3 fixing screws and remove the plate Remove the o ring...
Page 160: ...136 Remove the cardboard Completion of TMP mounting...
Page 161: ...137 12 Fix the flange to outlet Place the flange on the outlet Do not forget to place o ring...
Page 163: ...139 Slide fast the IS CTRL A board down and fix it 4 screws Connect the two connectors...
Page 164: ...140 14 Place the covers Place the front cover Place the front door...