ZEBV-5020C
3.
Detailed explanation of CI and NCI
Model
There are three types of QP2010 dependent on the types of ionization.
Ver. 2.1 of the GCMSsolution is required to use the CI, NCI, and DI option.
Description
Notes
Part Number
Remarks
GCMS-QP2010,
EI
without RP
The rotary pump RP1 must be
obtained separately. Order the
GCMSsolution Ver. 2.1 system
package together.
225-10040-xx
-92: 115V
-34: 230V*
C7
*CE marked
GCMS-QP2010c,
EI and CI
without RP
The rotary pump RP1 must be
obtained separately. Order the
GCMSsolution Ver. 2.1 system
package together. Ver. 2.1 of the
GCMSsolution is required to use the
CI.
225-10041-xx
-92: 115V
-34: 230V*
C7
*CE marked
GCMS-QP2010nc,
EI, CI, and NCI
without RP
The rotary pump RP1 must be
obtained separately. Order the
GCMSsolution Ver. 2.1 system
package together. Ver. 2.1 of the
GCMSsolution is required to use the
CI and NCI.
225-10042-xx
-92: 115V
-34: 230V*
C7
*CE marked
GCMSsolution Ver. 2.1
system package
Order this system package together
with GCMS-QP2010 main frame.
Ver. 2.1 of the GCMSsolution is
required to use the CI, NCI, and DI
option.
225-06628-92 C7
Major different points from 5050A
EI and CI (SEI, SCI) can be simulated using the NCI ion source (for the NCI equipped model
only)
Simulated CI (SCI) mode
in NCI ion source is added.
Ionization mode can be selected from SEI, SCI and NCI by indication from PC in NCI ion
source.
Note : SEI in CI ion source is not available.
Selection of reagent gases
QP2010 has
two inlet lines
for reagent gas.
Selection of reagent gas
is available from two
types of gas by switching the inlet line from PC.
RP2 is unnecessary for CI and NCI
The rotary pump, which is a backing pump for TMPs, evacuates the reagent gas line at the
same time.
Specification
[1] CI
1. Sensitivity
100pg benzophenone
Quasi-molecular ion S/N>150 at M/Z183 (RMS)
Reagent gas: methane, Scan range: 100-250u, Interval: 0.6sec
2. Reagent gas
Methane, isobutane, ammonia
3. Ion source
Ion source box exchange type
4. Reagent gas
introduction system
2 line system
5. Auto-tuning
Capable
6. Max. column flow
rate (He)
5mL/min (restricted compared to EI to allow introduction of the
reagent gas)
Summary of Contents for GCMS-QP2010 Series
Page 1: ...1 225 00612A Gas Chromatograph Mass Spectrometer GCMS QP2010 Service Manual...
Page 27: ...27...
Page 71: ...71 Wiring Diagrams 1 Wiring diagram Overall Wiring Diagrams...
Page 72: ...72 Wiring Diagrams 2 Wiring diagram MAIN POWER ASSY...
Page 73: ...73 Unit Operation Checks 3 Wiring diagram Details...
Page 76: ...76 MAIN POWER ASSY Block Diagram Parts inside the broken box are in the MAIN POWER ASSY...
Page 79: ...79 DC POWER A Block Diagram...
Page 84: ...84 MAIN CTRL A Block Diagram...
Page 85: ...85...
Page 90: ...90 IS CTRL A Block Diagram...
Page 94: ...94 HV PCB ASSY block diagram 1 UNNCI case of model EI CI 2 NCI case of model NCI...
Page 98: ...98 IG CTRL A PCB Block Diagram...
Page 109: ...109 DI CTRL A block diagram...
Page 150: ...126 Do not remove the 2 screws on the right side Remove the 2 screws on the left side...
Page 151: ...127 Remove cardboard Move the TW300 left by holding in both hands...
Page 155: ...131 Remove the cap of outlet Loosen the 3 fixing screws and remove the plate Remove the o ring...
Page 160: ...136 Remove the cardboard Completion of TMP mounting...
Page 161: ...137 12 Fix the flange to outlet Place the flange on the outlet Do not forget to place o ring...
Page 163: ...139 Slide fast the IS CTRL A board down and fix it 4 screws Connect the two connectors...
Page 164: ...140 14 Place the covers Place the front cover Place the front door...