11
Day 1
3. Environment Settings, Start-up
Turn on the instrument power supplies in the following sequence:
c
pretreatment unit
d
GC
e
MS
f
printer and monitor
g
PC
When the power is turned on for the first time after connecting the MS and PC, the MS driver
is installed automatically. Install the driver in accordance with section 8.8.3
Installing the
MS driver
of the System User Guide.
Outline Procedure (See the System User Guide for more details.)
Press [Next (N)].
Display the list of the known drivers of the device.
Select “Another Device”.
Press [Next (N)].
Press [Use Disk].
Insert the CD-ROM
Enter "E:¥Driver" and press [Enter].
Select “Shimadzu GCMS-QP2010”.
Press [Next (N)].
Press [Next (N)].
Environment settings
Set up the software environment, as described in section
3.3 System Configuration
of the
Operation Guide.
Start-up
Start evacuation, as described in section
4.1 Starting and Stopping the Vacuum System
of
the Operation Guide.
Hint
Summary of Contents for GCMS-QP2010 Series
Page 1: ...1 225 00612A Gas Chromatograph Mass Spectrometer GCMS QP2010 Service Manual...
Page 27: ...27...
Page 71: ...71 Wiring Diagrams 1 Wiring diagram Overall Wiring Diagrams...
Page 72: ...72 Wiring Diagrams 2 Wiring diagram MAIN POWER ASSY...
Page 73: ...73 Unit Operation Checks 3 Wiring diagram Details...
Page 76: ...76 MAIN POWER ASSY Block Diagram Parts inside the broken box are in the MAIN POWER ASSY...
Page 79: ...79 DC POWER A Block Diagram...
Page 84: ...84 MAIN CTRL A Block Diagram...
Page 85: ...85...
Page 90: ...90 IS CTRL A Block Diagram...
Page 94: ...94 HV PCB ASSY block diagram 1 UNNCI case of model EI CI 2 NCI case of model NCI...
Page 98: ...98 IG CTRL A PCB Block Diagram...
Page 109: ...109 DI CTRL A block diagram...
Page 150: ...126 Do not remove the 2 screws on the right side Remove the 2 screws on the left side...
Page 151: ...127 Remove cardboard Move the TW300 left by holding in both hands...
Page 155: ...131 Remove the cap of outlet Loosen the 3 fixing screws and remove the plate Remove the o ring...
Page 160: ...136 Remove the cardboard Completion of TMP mounting...
Page 161: ...137 12 Fix the flange to outlet Place the flange on the outlet Do not forget to place o ring...
Page 163: ...139 Slide fast the IS CTRL A board down and fix it 4 screws Connect the two connectors...
Page 164: ...140 14 Place the covers Place the front cover Place the front door...