114
2. Adjustment in the high mass-number range
This instrument incorporates an auto-tuning function that conducts mass-number calibration
at m/z614. Mass-number displacement is rare in the normal range of operation (up to
approximately m/z700.)
However, displacement from the calibration at m/z is possible in the high mass-number
range.
Ensuring accurate calibration in the range m/z700 to m/z1000 requires manual calibration
by injecting a triazine (Tris(perfluoroheptyi)-S-triazine) high-mass-number standard sample
into the column.
Procedure
Instrument conditions
Sample: triazine 1% solution in acetone, ultrasonically mixed
INJ: 250
M
C
COL: 50
M
C (1 min.)-5
M
C /min-100
M
C (1min)-30
M
C /min-250
M
C (10min)
I/F: 250
M
C
Ion source: 200
M
C
Column head pressure: 100kPa
Total flow rate: 30mL/minute
Splitless injection (1 min. sampling)
Triazine flow time is approximately 3 minutes between minute 11 and minute 15
c
Take 1
µ
L triazine sample into a syringe and inject it into INJ. Wipe any sample from
the syringe tip before injecting the sample.
d
Monitor m/z771, 886, and 1016 on the Peak Monitor window. Leave the filament off
for the first five minutes after sample injection, turn the filament on, and wait for the
peaks to appear.
e
The triazine peak can be seen for the three minutes between minute 11 and minute 15.
Check the mass number (within
C
0.3 mass) and resolution (FWHM resolution 0.1M
max.). Conduct manual compensation, if necessary.
f
After making this check, increase the column temperature to 250
M
C to drive out the
triazine. Ten minutes at 250
M
C is insufficient, conduct this process for 1 hour at
250
M
C.
Caution) Traces of the injected triazine may remain in the sample lines and
hinder subsequent analysis. Allow the column to age overnight at
250
M
C before conducting ultramicroanalysis.
Summary of Contents for GCMS-QP2010 Series
Page 1: ...1 225 00612A Gas Chromatograph Mass Spectrometer GCMS QP2010 Service Manual...
Page 27: ...27...
Page 71: ...71 Wiring Diagrams 1 Wiring diagram Overall Wiring Diagrams...
Page 72: ...72 Wiring Diagrams 2 Wiring diagram MAIN POWER ASSY...
Page 73: ...73 Unit Operation Checks 3 Wiring diagram Details...
Page 76: ...76 MAIN POWER ASSY Block Diagram Parts inside the broken box are in the MAIN POWER ASSY...
Page 79: ...79 DC POWER A Block Diagram...
Page 84: ...84 MAIN CTRL A Block Diagram...
Page 85: ...85...
Page 90: ...90 IS CTRL A Block Diagram...
Page 94: ...94 HV PCB ASSY block diagram 1 UNNCI case of model EI CI 2 NCI case of model NCI...
Page 98: ...98 IG CTRL A PCB Block Diagram...
Page 109: ...109 DI CTRL A block diagram...
Page 150: ...126 Do not remove the 2 screws on the right side Remove the 2 screws on the left side...
Page 151: ...127 Remove cardboard Move the TW300 left by holding in both hands...
Page 155: ...131 Remove the cap of outlet Loosen the 3 fixing screws and remove the plate Remove the o ring...
Page 160: ...136 Remove the cardboard Completion of TMP mounting...
Page 161: ...137 12 Fix the flange to outlet Place the flange on the outlet Do not forget to place o ring...
Page 163: ...139 Slide fast the IS CTRL A board down and fix it 4 screws Connect the two connectors...
Page 164: ...140 14 Place the covers Place the front cover Place the front door...