36
Description of Parts
8. Turbo Molecular Pump
Two turbo molecular pumps (TMP), one for the ion-source unit exhaust (TW-300) and one
for the analyzer exhaust (TW-70), are connected to a single rotary pump (E2M1.5).
Removing the Turbo Molecular Pumps
1. Remove IS Cont PCB. Remove the IG tube.
2. Disconnect the TMP exhaust unit from the rotary pump. Loosen and remove the
clamp from the KF flange. The TMP1 flange is integral with the TMP body and can be
difficult to remove.
3. Disconnect the connector (power) from the power supply.
4. Remove the TMP fixing screws and remove the TMP.
5. Lower TMP1 to the top of the TMP power supply, slide it to the left (as seen from the
front), and remove it from the instrument.
6. Reverse this procedure to install the TMP.
Caution Ensure that no dust or dirt sticks to the O-ring at the top of the TMP.
(Purge with dry gas (He, N
2
) during installation.)
RP connection (TMP exhaust unit)
TMP fixing screws
(4 positions)
TMP1
(TW300)
TMP2
(TW70)
Summary of Contents for GCMS-QP2010 Series
Page 1: ...1 225 00612A Gas Chromatograph Mass Spectrometer GCMS QP2010 Service Manual...
Page 27: ...27...
Page 71: ...71 Wiring Diagrams 1 Wiring diagram Overall Wiring Diagrams...
Page 72: ...72 Wiring Diagrams 2 Wiring diagram MAIN POWER ASSY...
Page 73: ...73 Unit Operation Checks 3 Wiring diagram Details...
Page 76: ...76 MAIN POWER ASSY Block Diagram Parts inside the broken box are in the MAIN POWER ASSY...
Page 79: ...79 DC POWER A Block Diagram...
Page 84: ...84 MAIN CTRL A Block Diagram...
Page 85: ...85...
Page 90: ...90 IS CTRL A Block Diagram...
Page 94: ...94 HV PCB ASSY block diagram 1 UNNCI case of model EI CI 2 NCI case of model NCI...
Page 98: ...98 IG CTRL A PCB Block Diagram...
Page 109: ...109 DI CTRL A block diagram...
Page 150: ...126 Do not remove the 2 screws on the right side Remove the 2 screws on the left side...
Page 151: ...127 Remove cardboard Move the TW300 left by holding in both hands...
Page 155: ...131 Remove the cap of outlet Loosen the 3 fixing screws and remove the plate Remove the o ring...
Page 160: ...136 Remove the cardboard Completion of TMP mounting...
Page 161: ...137 12 Fix the flange to outlet Place the flange on the outlet Do not forget to place o ring...
Page 163: ...139 Slide fast the IS CTRL A board down and fix it 4 screws Connect the two connectors...
Page 164: ...140 14 Place the covers Place the front cover Place the front door...