2
Introduction
This Service Manual is intended for service personnel authorized by Shimadzu Corporation.
It describes maintenance of the GCMS-QP2010 instrument.
Untrained personnel must not attempt maintenance work on the instrument. No
responsibility is accepted for work conducted by untrained personnel.
Safety Precautions
Before starting work, read and fully understand the Danger, Warning, Caution, and Notes
safety items in the yellow page at the beginning of the GCMS-QP2010 System User Guide
(225-00822).
The GCMS-QP2010 contains hot parts, such as the GC2010 heater unit, interface heater
unit, and ion-source heater unit, a high-voltage power supply for the detector unit, and
high-voltage circuits such as the vacuum-measurement circuit. Take sufficient care when
handling high-temperature and high-voltage units.
Disclaimers
(1) The copyright of this document belongs to Shimadzu Corporation, Ltd. The copying or
reproduction of part or all of this manual is prohibited without the permission of Shimadzu
Corporation.
(2) Some parts in this Service Manual may differ from the actual instrument. Due to
product modifications, we reserve the right to change this manual without notice.
(3) This Service Manual is considered complete at the time of production. Any errors or
omissions discovered may not be remedied immediately.
Copyright (C) 2001 SHIMADZU CORPORATION
Summary of Contents for GCMS-QP2010 Series
Page 1: ...1 225 00612A Gas Chromatograph Mass Spectrometer GCMS QP2010 Service Manual...
Page 27: ...27...
Page 71: ...71 Wiring Diagrams 1 Wiring diagram Overall Wiring Diagrams...
Page 72: ...72 Wiring Diagrams 2 Wiring diagram MAIN POWER ASSY...
Page 73: ...73 Unit Operation Checks 3 Wiring diagram Details...
Page 76: ...76 MAIN POWER ASSY Block Diagram Parts inside the broken box are in the MAIN POWER ASSY...
Page 79: ...79 DC POWER A Block Diagram...
Page 84: ...84 MAIN CTRL A Block Diagram...
Page 85: ...85...
Page 90: ...90 IS CTRL A Block Diagram...
Page 94: ...94 HV PCB ASSY block diagram 1 UNNCI case of model EI CI 2 NCI case of model NCI...
Page 98: ...98 IG CTRL A PCB Block Diagram...
Page 109: ...109 DI CTRL A block diagram...
Page 150: ...126 Do not remove the 2 screws on the right side Remove the 2 screws on the left side...
Page 151: ...127 Remove cardboard Move the TW300 left by holding in both hands...
Page 155: ...131 Remove the cap of outlet Loosen the 3 fixing screws and remove the plate Remove the o ring...
Page 160: ...136 Remove the cardboard Completion of TMP mounting...
Page 161: ...137 12 Fix the flange to outlet Place the flange on the outlet Do not forget to place o ring...
Page 163: ...139 Slide fast the IS CTRL A board down and fix it 4 screws Connect the two connectors...
Page 164: ...140 14 Place the covers Place the front cover Place the front door...