ZEBV-5020C
Installation example (The dimensions of the PC and printer may vary according to the model.)
See below for the cable and pipe lengths.
MS to rotary pump
vacuum pipe
approx. 1.5m
power
cable
approx.
1.5m
PC
to
MS
cable
approx.
2m
(Note
1)
Reference
number
Description
Dimensions (mm)
Weight (kg)
W
D
H
1 GC 515
530
440
30
2 MS 345
494
390
44.5
3 Monitor 390
189
364
4.7
4 Keyboard 454
173
54 1
5 PC 171
326
323
6.5
6 Printer 387
283
274 7
7
Oil rotary pump
125
323
189
10
Note 1 A 4.5m cable is also available to install the PC and MS further apart.
Order it if required.
Cat. No.
Description
070-60311-03 Cable
AXJ231132
(4.5m)
4. Power requirements
Caution
Provide the instrument with a separate power supply protected by an earth
leakage breaker. It must not share a power supply with any other equipment.
Caution
To prevent electric shocks, connect to earth with a grounding resistance not
exceeding 100
Ω
.
Note
Although the instrument is designed to operate normally with
±
10% AC line
voltage fluctuations, including accumulated rapid noise, to ensure performance
specifications are met, the voltage fluctuations (including noise) should not
exceed
±
5%. The power supply equipment must be improved if the voltage
fluctuations exceed
±
10%. Frequency fluctuations must not exceed
±
0.5Hz.
The power supply requirements for each unit are listed below.
* 115V instrument
Unit
Voltage
Capacity
Cable length
Terminal type
GC 115VAC,
1800VA 2.0m Fig.
1
Summary of Contents for GCMS-QP2010 Series
Page 1: ...1 225 00612A Gas Chromatograph Mass Spectrometer GCMS QP2010 Service Manual...
Page 27: ...27...
Page 71: ...71 Wiring Diagrams 1 Wiring diagram Overall Wiring Diagrams...
Page 72: ...72 Wiring Diagrams 2 Wiring diagram MAIN POWER ASSY...
Page 73: ...73 Unit Operation Checks 3 Wiring diagram Details...
Page 76: ...76 MAIN POWER ASSY Block Diagram Parts inside the broken box are in the MAIN POWER ASSY...
Page 79: ...79 DC POWER A Block Diagram...
Page 84: ...84 MAIN CTRL A Block Diagram...
Page 85: ...85...
Page 90: ...90 IS CTRL A Block Diagram...
Page 94: ...94 HV PCB ASSY block diagram 1 UNNCI case of model EI CI 2 NCI case of model NCI...
Page 98: ...98 IG CTRL A PCB Block Diagram...
Page 109: ...109 DI CTRL A block diagram...
Page 150: ...126 Do not remove the 2 screws on the right side Remove the 2 screws on the left side...
Page 151: ...127 Remove cardboard Move the TW300 left by holding in both hands...
Page 155: ...131 Remove the cap of outlet Loosen the 3 fixing screws and remove the plate Remove the o ring...
Page 160: ...136 Remove the cardboard Completion of TMP mounting...
Page 161: ...137 12 Fix the flange to outlet Place the flange on the outlet Do not forget to place o ring...
Page 163: ...139 Slide fast the IS CTRL A board down and fix it 4 screws Connect the two connectors...