16
Day 2
6. Qualitative Analysis (SCAN Mode)
Describe the method of qualitative analysis using a standard sample, such as OFN
100pg/
µ
L.
Operation Description
Explain the file types described in Section
2.9 Common Operations
of the System User
Guide and how to use Data Explorer, and create a new project folder.
Explain the analysis method and data analysis method in accordance with section
6
Qualitative Analysis (SCAN Mode)
of the Operation Guide.
Points to Explain to the Customer
How to use Explorer
How to use the Method Wizard and how to check the methods created
Carrier-gas pressure setting range (column flow to 15mL/minute)
Meaning of "solvent elution time"
Standard values of each parameter
How to save method files
How to register samples
How to run the data-processing program
How to expand TIC
How to eliminate background noise
How to display mass-chromatographs
How to search libraries
How to register spectral processing tables
Conduct an S/N check, if time is available.
See section
8. S/N Measurements
for details about the measurement conditions.
Summary of Contents for GCMS-QP2010 Series
Page 1: ...1 225 00612A Gas Chromatograph Mass Spectrometer GCMS QP2010 Service Manual...
Page 27: ...27...
Page 71: ...71 Wiring Diagrams 1 Wiring diagram Overall Wiring Diagrams...
Page 72: ...72 Wiring Diagrams 2 Wiring diagram MAIN POWER ASSY...
Page 73: ...73 Unit Operation Checks 3 Wiring diagram Details...
Page 76: ...76 MAIN POWER ASSY Block Diagram Parts inside the broken box are in the MAIN POWER ASSY...
Page 79: ...79 DC POWER A Block Diagram...
Page 84: ...84 MAIN CTRL A Block Diagram...
Page 85: ...85...
Page 90: ...90 IS CTRL A Block Diagram...
Page 94: ...94 HV PCB ASSY block diagram 1 UNNCI case of model EI CI 2 NCI case of model NCI...
Page 98: ...98 IG CTRL A PCB Block Diagram...
Page 109: ...109 DI CTRL A block diagram...
Page 150: ...126 Do not remove the 2 screws on the right side Remove the 2 screws on the left side...
Page 151: ...127 Remove cardboard Move the TW300 left by holding in both hands...
Page 155: ...131 Remove the cap of outlet Loosen the 3 fixing screws and remove the plate Remove the o ring...
Page 160: ...136 Remove the cardboard Completion of TMP mounting...
Page 161: ...137 12 Fix the flange to outlet Place the flange on the outlet Do not forget to place o ring...
Page 163: ...139 Slide fast the IS CTRL A board down and fix it 4 screws Connect the two connectors...
Page 164: ...140 14 Place the covers Place the front cover Place the front door...