58
Problems when starting evacuation
5. No ready status after automatic start
Check the GC settings.
□
Is an appropriate carrier gas pressure set at the GC?
An inappropriate carrier gas supply can result in no pressure increase at the column
inlet, so that the automatic startup sequence does not begin (and the rotary pump
does not operate). Check that the set carrier gas conditions are suitable for the
column.
Check the rotary pump.
□
Is the rotary pump rotating?
If not, check the wiring.
If the wiring is correct, see section
4.4 Main Power Assy
.
Check for leaks.
□
Is the unit front door completely closed?
Occasionally, dust can cause minute leaks. Check that no dust is sticking to the
O-ring contact faces before closing the front door.
□
Is there any leakage from the interface column connector?
Leakage can occur from the interface unit after repeated analysis with heating.
Tighten the connector slightly (about 1/6 turn).
Caution: Fully tightening the connector with the wrench supplied could damage the
screw threads. If slight re-tightening does not stop the leakage, replace the ferrule.
□
Is the column broken?
It is impossible to maintain a vacuum if the column breaks near the interface unit. If
the column is scratched during maintenance, it tends to break at the scratch.
□
If the analysis tube was opened during maintenance, is there any leakage from it?
If a normal vacuum could be maintained before maintenance but not after it, initially
check the positions where maintenance work were conducted and the interface
column connector (the temperature has been dropped).
Summary of Contents for GCMS-QP2010 Series
Page 1: ...1 225 00612A Gas Chromatograph Mass Spectrometer GCMS QP2010 Service Manual...
Page 27: ...27...
Page 71: ...71 Wiring Diagrams 1 Wiring diagram Overall Wiring Diagrams...
Page 72: ...72 Wiring Diagrams 2 Wiring diagram MAIN POWER ASSY...
Page 73: ...73 Unit Operation Checks 3 Wiring diagram Details...
Page 76: ...76 MAIN POWER ASSY Block Diagram Parts inside the broken box are in the MAIN POWER ASSY...
Page 79: ...79 DC POWER A Block Diagram...
Page 84: ...84 MAIN CTRL A Block Diagram...
Page 85: ...85...
Page 90: ...90 IS CTRL A Block Diagram...
Page 94: ...94 HV PCB ASSY block diagram 1 UNNCI case of model EI CI 2 NCI case of model NCI...
Page 98: ...98 IG CTRL A PCB Block Diagram...
Page 109: ...109 DI CTRL A block diagram...
Page 150: ...126 Do not remove the 2 screws on the right side Remove the 2 screws on the left side...
Page 151: ...127 Remove cardboard Move the TW300 left by holding in both hands...
Page 155: ...131 Remove the cap of outlet Loosen the 3 fixing screws and remove the plate Remove the o ring...
Page 160: ...136 Remove the cardboard Completion of TMP mounting...
Page 161: ...137 12 Fix the flange to outlet Place the flange on the outlet Do not forget to place o ring...
Page 163: ...139 Slide fast the IS CTRL A board down and fix it 4 screws Connect the two connectors...
Page 164: ...140 14 Place the covers Place the front cover Place the front door...