231
16 Special Functions
16.4
16
.
GC-2014
16.4
Pre-Run Program
The Pre-Run program controls flow controllers and relays before the analysis starts.
16.4.1
Screen description
Select “4. PRE-RUN” from the [FUNC] key screen to display the Pre-Run screen shown in
Fig. 16.4.1. A Pre-Run program is executed after the [START] key is pressed, but before an
analysis.
While the Pre-Run program is running, the elapsed time is displayed in the “Time” field on
the [MONIT] key screen.
Program steps do not need to be entered in order. After you finish editing the Pre-Run
program, programs are automatically sorted.
More than one step can be set for the same time. The specified actions are executed at the
same time.
16.4.2
PF menu
Fig. 16.4.1
Pre-Run program main screen
PF menu
Description
Reference
section
New
Creates a new Pre-Run program.
16.4.3
Del Line
Deletes the program at the current cursor position.
――
Chng Line
Changes over the screen of Pre-Run program every line.
――
Prints the program to the Chromatopac.
――
Summary of Contents for GC-2014
Page 4: ...This page is intentionally left blank ...
Page 12: ...Contents viii GC 2014 This page is intentionally left blank ...
Page 28: ...1 Installation 1 3 Gas Supply Plumbing 16 GC 2014 This page is intentionally left blank ...
Page 78: ...5 Analysis 5 2 Notes for Analysis 66 GC 2014 This page is intentionally left blank ...
Page 90: ...6 Basic Key Operation 6 4 Getting Help 78 GC 2014 This page is intentionally left blank ...
Page 282: ...16 Special Functions 16 10 ROM Version No 270 GC 2014 This page is intentionally left blank ...
Page 318: ...306 GC 2014 This page is intentionally left blank ...