14 Diagnosis
14.4 Coolant Consumption Counter
191
GC-2014
14.4.2
Parameter list
USE COUNTER
Selection: Yes/No, Default : Yes
CONSUMPTION
Default: 0 min
The total time that the CRG is “On” is displayed.
Refer to
“15.5 Setting the CRG Parameters”
SETTING TO WARN
Range: 0
−
9999 min, Default : 100 min
The coolant consumption time depends on the gas cylinder volume and the number of analyses which
requier CRG. Set the warning time according to the analysis conditions.
14.4.3
PF menu
PF menu
Description
Reference
section
Reset
Resets the Use counter.
――
NOTE