4 Installing Capillary Columns and Setting Analytical Flow Lines
4.2 Capillary Analysis Using the Dual FID (When a detector adapter with purge is used)
51
GC-2014
3.
Set the makeup gas flow rate.
(1) Press the [OPTION] key.
(2) Set AMC.L when the detector adopter is connected to the left (L) side of the dual INJ
and set AMC.R when it is connected to the right (R) side.
Start control and set the makeup gas flow rate and gas type.
4.
Set the dual FID.
(1) Press the [DET] key.
(2) Set “Control mode” as described below.
Set to “Single L” when the column is connected to the left (L) side of the dual FID and “Single
R” when it is connected to the right (R) side.