18 Error Messages
18.1 Error Messages
301
GC-2014
18.1.9
Warning messages
These messages appear when the time or the count exceeds the set value. Replace the
subject component and reset the count on the menu of the DIAG key screen.
Code 9002 warning requires component replacement and counter reset by a serviceperson.
Contact your Shimadzu representative.
This message appears when the system was started before it was ready. Normally, do not
start analysis until the system is ready.
If this message appears when all the parameters including temperature and flow rate are
ready, check the ready setting for unused components and check the equilibration time.
This message appears when ignition sequence was re-attempted because the ignition
failed. This does not stop the execution of analysis. If this message appears frequently,
check “DETECTOR IGNITE” and “DETECTOR IGNITION” by the standard diagnosis
procedure. Also check for gas leaks and verify the gas flow rates. If set values are correct,
contact your Shimadzu representative.
These messages appear when the sensor use time exceeds the preset value. Component
replacement and reset by a serviceperson are required. Contact your Shimadzu representative.
Code
Message
CS
9000
COL CRG use time is over
Yes
9001
INJ2 CRG use time is over
Yes
9002
Fan use time is over
Yes
9004
CAR1 septum counter is over
No
9005
CAR2 septum counter is over
No
9006
CAR3 septum counter is over
No
9007
CAR1 insert counter is over
No
9008
CAR2 insert counter is over
No
9009
CAR3 insert counter is over
No
Code
Message
CS
9010
System is not ready
No
Code
Message
CS
9011
Ignition finished (retried)
No
Code
Message
CS
9012
COL sensor use time is over
Yes
9013
INJ1 sensor use time is over
Yes
9014
DET1 sensor use time is over
Yes
9015
INJ2 sensor use time is over
Yes
9016
DET2 sensor use time is over
Yes
9017
AUX3 sensor use time is over
Yes
9018
AUX4 sensor use time is over
Yes
9019
AUX5 sensor use time is over
Yes
Summary of Contents for GC-2014
Page 4: ...This page is intentionally left blank ...
Page 12: ...Contents viii GC 2014 This page is intentionally left blank ...
Page 28: ...1 Installation 1 3 Gas Supply Plumbing 16 GC 2014 This page is intentionally left blank ...
Page 78: ...5 Analysis 5 2 Notes for Analysis 66 GC 2014 This page is intentionally left blank ...
Page 90: ...6 Basic Key Operation 6 4 Getting Help 78 GC 2014 This page is intentionally left blank ...
Page 282: ...16 Special Functions 16 10 ROM Version No 270 GC 2014 This page is intentionally left blank ...
Page 318: ...306 GC 2014 This page is intentionally left blank ...