16 Special Functions
16.6 GC Configuration
253
GC-2014
16.6.11.2
Parameter list
LANGUAGE
Selection: Alt./English, Default: English
Switch between Japanese (Alt.) and English screen languages.
BACK LIGHT AUTO OFF
Range: 0
−
9999 (sec)., Default: 0 (sec)
Set the period of time when the backlight LCD display automatically turns off.
PRESSURE UNIT
Selection: kPa/bar/psi., Default: kPa
Set the pressure units for display.
In the case that the presssure unit of “bar” or “psi” are used, pay attention to the following.
•
The range of the primary pressure is not displayed in the value of “bar”, “psi”. Set up it
with the value that consults the following table and transformed into “kPa”.
•
The pressure axis on the screen after pressing [MONIT] key is displayed in the value of
“kPa” regardless of the “pressure unit” sets.
•
The setting value, measured value of the primary pressure of the diagnosis result of the
standard diagnosis is displayed in the value of “kPa” regardless of the “pressure unit”
sets.
•
The display, printing of each log file is displayed in the value of “kPa” regardless of the
“pressure unit” sets.
BEEP VOLUME
Selection: Off/Low/Mid/Hig, Default: Low
BEEP TONE
Selection: Lo/Hi, Default: Hi
Use a different tone for different GCs to be able to discern which GC has a problem.
ATMOSPHERIC COMPENSATION
Selection: On/Off, Default: Off
Select “On” to compensate for higher altitudes when in linear velocity mode.
ZERO AT READY
Selection: On/Off, Default: On
Set “On” for, the gas chromatograph to automatically perform zero adjustment when the
GC is ready.
POLARITY IN READY
Selection: Open/Close, Default: Open
Set the polarity of GC ready signal.
INJ1/INJ2 PRIMARY PRESS
Selection: 300
−
500 kPa/500
−
900 kPa/900
−
980 kPa, Default: 500
−
900 kPa
Select the pressure range of the carrier gas cylinder.
If the wrong range is selected, error messages are not displayed correctly.
Example: If the gas cylinder pressure supplied to the GC-2014 is 700 kPa, select “500
−
900 kPa”. This is not displayed when the carrier gas flow controller is DAFC.
Range (kPa = 1)
Increase quantity
kPa
1
0.1
bar
1/100
0.01
psi
1/6.895
0.1
kPa
bar
psi
300 - 500
3.00 - 5.00
43.5 - 72.5
500 - 900
5.00 - 9.00
72.5 - 130.5
900 - 980
9.00 - 9.80
130.5 - 142.1
NOTE
Summary of Contents for GC-2014
Page 4: ...This page is intentionally left blank ...
Page 12: ...Contents viii GC 2014 This page is intentionally left blank ...
Page 28: ...1 Installation 1 3 Gas Supply Plumbing 16 GC 2014 This page is intentionally left blank ...
Page 78: ...5 Analysis 5 2 Notes for Analysis 66 GC 2014 This page is intentionally left blank ...
Page 90: ...6 Basic Key Operation 6 4 Getting Help 78 GC 2014 This page is intentionally left blank ...
Page 282: ...16 Special Functions 16 10 ROM Version No 270 GC 2014 This page is intentionally left blank ...
Page 318: ...306 GC 2014 This page is intentionally left blank ...