15 Optional Devices
15.5 Setting the CRG Parameters
210
GC-2014
15.5.2
Parameter list
COLUMN OVEN
VALVE
Selection: On/Off, Default: Off
CRG
Selection: Use/Not Use, Default: Not Use
When the CRG is connected to the column oven, the column oven temperature
control range is extended to -99.0
−
+400.0 °C from (room tempe 4 °C)
−
+400.0 °C.
When using the CRG, set “VALVE” to “On” and “CRG” to “Use”.
When using valve control only, set “VALVE” to “On” and set “CRG” to “Not Use”.
INJ 2
VALVE
Selection: On/Off, Default: Off
CRG
Selection: Use/ Not Use, Default: Not Use
When the PTV is installed on INJ2, the injection port temperature control range is
extended to -99.0
−
+400.0 °C from 0.0
−
400.0 °C
When using the CRG, set “VALVE” to “On” and “CRG” to “Use”.
When using valve control only, set “VALVE” to “On” and set “CRG” to “Not Use”.
15.5.3
PF menu
PF menu
Description
Reference
section
Next
Toggles the screen among AOC parameters
→
AUX temperature
→
AUX
APC
→
AUX AMC
→
CRG screens in this order.
――
Summary of Contents for GC-2014
Page 4: ...This page is intentionally left blank ...
Page 12: ...Contents viii GC 2014 This page is intentionally left blank ...
Page 28: ...1 Installation 1 3 Gas Supply Plumbing 16 GC 2014 This page is intentionally left blank ...
Page 78: ...5 Analysis 5 2 Notes for Analysis 66 GC 2014 This page is intentionally left blank ...
Page 90: ...6 Basic Key Operation 6 4 Getting Help 78 GC 2014 This page is intentionally left blank ...
Page 282: ...16 Special Functions 16 10 ROM Version No 270 GC 2014 This page is intentionally left blank ...
Page 318: ...306 GC 2014 This page is intentionally left blank ...