2 Before Use
2.2 Outputting Analog Signals to the Chromatopac
21
GC-2014
■
Setting analog signal output
1.
Set a line
(1) Press the [SET] key then press the
[PF2] key (Line Config).
(2) Specify an injection port and detector
to which a column is connected in the
same line.
The screen on the right shows an
example where a column is connected
to the dual INJ and dual FID.
Without specifying a detector in a line, the screen to set analog signal output below does not appear.
2.
Set analog signal output.
(1) Press the [DET] key.
(2) The outlined part is a parameter for
analog signal output for all detectors.
Set the parameter following the
description below.
For “Background signal save/
compensation” and “Detector signal
subtraction,” refer to “Chapter 13
Detector [DET].”
“Signal Output Port”
Specify a connector number to output
analog signals.
(ANALOG OUT 1 = Ch1, ANALOG OUT 2 = Ch2)
Channel number of a detector specified at
the last is effective. (If TCD signal was
outputted to Ch1 formerly and FID signal is
newly specified to be outputted to Ch1 as
shown in the screen on the right, TCD
signal becomes OFF automatically.)
NOTE
Summary of Contents for GC-2014
Page 4: ...This page is intentionally left blank ...
Page 12: ...Contents viii GC 2014 This page is intentionally left blank ...
Page 28: ...1 Installation 1 3 Gas Supply Plumbing 16 GC 2014 This page is intentionally left blank ...
Page 78: ...5 Analysis 5 2 Notes for Analysis 66 GC 2014 This page is intentionally left blank ...
Page 90: ...6 Basic Key Operation 6 4 Getting Help 78 GC 2014 This page is intentionally left blank ...
Page 282: ...16 Special Functions 16 10 ROM Version No 270 GC 2014 This page is intentionally left blank ...
Page 318: ...306 GC 2014 This page is intentionally left blank ...