4 Installing Capillary Columns and Setting Analytical Flow Lines
4.4 Capillary TCD Analysis
55
GC-2014
■
Setting analytical flow lines
Setting analytical flow lines is described below.
For procedures to install columns and input analytical conditions, refer to the operation
manual.
1.
Change the dual AFC configuration.
(1) Select “6. GC CONFIGURATION” on the [FUNC] key screen and then select “9. Other
configuration.”
(2) Specify “AMC.LR” at “DAFC unit.”
2.
Set a line.
(1) Press the [SET] key and then press the [PF2] key (Line Config).
(2) Set the SPL and dual TCD (DTCD) on a same line.
LINE1 is specified in the example below. Lines 1 through 4 can be selected.
If the dual AFC is specified for AMC.LR, the indication of DINJ disappears.
NOTE