![Shimadzu GC-2014 Instruction Manual Download Page 288](http://html1.mh-extra.com/html/shimadzu/gc-2014/gc-2014_instruction-manual_1256914288.webp)
17 Printing
17.2 AOC commands
276
GC-2014
1: Without autosampler, short rack: 1
−
6
Without autosampler, long rack: 1
−
12
With autosampler: 1
−
150
Command names marked with “ * ” cannot be set on the GC screen.
SAMU
Use of the sampler
Not use: 0, Use: 1
Not use: 0
SUBU
Use of the sub AOC
Not use: 0, Use: 1
Not use: 0
BARC
Bar Code Reader
Not use: 0, Use: 1
Not use: 0
SPMD
Distribution of sample for dual AOC
0
−
8
0
PAR1
Use of Same Param
No: 0, Yes: 1
No: 0
*GLPM
Setting of validation mode
0, 1
0
*GRPT
Number of times of GLP sample discharge
1
−
99
50
*GVOL
GLP sample discharge volume
1
−
80
20
*GPMP
Number of times of pumping after second
GLP
0
−
5
1
*ATSP
Automatic stop function
Off: 0, On: 1
On: 1
*ARSG
Ready signal polarity
Open: 0, Close: 1
Open: 0
TLET
Rack
Short: 0, Long: 1
Short: 0
TSEL
Rack position while sampler is used
0
−
2
1
Command
name
Item
Range
Default
NOTE
Summary of Contents for GC-2014
Page 4: ...This page is intentionally left blank ...
Page 12: ...Contents viii GC 2014 This page is intentionally left blank ...
Page 28: ...1 Installation 1 3 Gas Supply Plumbing 16 GC 2014 This page is intentionally left blank ...
Page 78: ...5 Analysis 5 2 Notes for Analysis 66 GC 2014 This page is intentionally left blank ...
Page 90: ...6 Basic Key Operation 6 4 Getting Help 78 GC 2014 This page is intentionally left blank ...
Page 282: ...16 Special Functions 16 10 ROM Version No 270 GC 2014 This page is intentionally left blank ...
Page 318: ...306 GC 2014 This page is intentionally left blank ...