263
16 Special Functions
16.8
16
.
GC-2014
16.8
Stopwatch
16.8.1
Screen description
Select “8. STOP WATCH” from the [FUNC] key screen to display the stopwatch screen
shown in Fig. 16.8.1.
The stopwatch can display elapsed time up to 99 : 99 : 99.9 in units of 0.1 second. When the
counted time exceeds 99 : 99 : 99.9, the stopwatch is reset and the time restants at 0.0
seconds.
The stopwatch can function even when other keys are pressed. However, once the stop key
is pressed, the stopwatch is reset to 0.0 seconds if another key is pressed.
16.8.2
PF menu
16.8.3
Timing with inverse measurement
When using a bubble film flow meter of V ml, obtain the flow rate using the reciprocal
number.
Fig. 16.8.1
Stop watch screen
PF menu
Description
Reference
section
Start
Starts timing.
If “Start” is pressed again when the timing has stopped, the stopwatch is
reset to 0.0 seconds.
――
Stop
Stops counting.
――
V x (Reciprocal number of measurement time) ml/min