![Shimadzu GC-2014 Instruction Manual Download Page 117](http://html1.mh-extra.com/html/shimadzu/gc-2014/gc-2014_instruction-manual_1256914117.webp)
9 Monitoring the GC
9.1 [MONIT] key Main Screen
105
GC-2014
9.1.2
Parameter list
RETENTION TIME (Rt)
Retention time is the length of time from compound injectio to detection.
The retention time is specific to each compound. Compounds are identified based on match-
ing the retention time to a standard compound retention time.
REMAIN TIME
The longest program time subtracted by the current retention time is displayed.
LONGEST PROGRAM TIME
The longest program time indicates the total time of the longest program when comparing
the temperature program, the pressure program, the flow rate program and the time
program.
9.1.3
PF menu
PF menu
Description
Reference
section
Temp Mon
Displays the current and set temperatures of column, injection port and
detector.
9.2
Flow Mon
Displays inlet pressure, total flow rate and purge flow rate for injection port,
and hydrogen flow rate, air flow rate and makeup gas flow rate for detector.
The display format depends on the typeof flow controller installed.
9.3
Zero Adj
Automatically moves the baseline to zero point.
9.4
Zero Free
Returns to the baseline level before zero point adjustment.
9.4
Up
Moves the baseline up by 100 µV from current level.
9.4
Down
Moves the baseline down by 100 µV from current level.
9.4
Chng Graph
Switches the graph display from chromatogram and column oven tempera-
ture program to chromatogram and pressure program. For direct injection
mode with flow control, the flow rate program is displayed instead of the
pressure program.
――
Chng Line
Displays the monitor screen for another configured analytical line. Switch
between monitored lines by pressing [MONIT] key from the monitor screen.
――
Summary of Contents for GC-2014
Page 4: ...This page is intentionally left blank ...
Page 12: ...Contents viii GC 2014 This page is intentionally left blank ...
Page 28: ...1 Installation 1 3 Gas Supply Plumbing 16 GC 2014 This page is intentionally left blank ...
Page 78: ...5 Analysis 5 2 Notes for Analysis 66 GC 2014 This page is intentionally left blank ...
Page 90: ...6 Basic Key Operation 6 4 Getting Help 78 GC 2014 This page is intentionally left blank ...
Page 282: ...16 Special Functions 16 10 ROM Version No 270 GC 2014 This page is intentionally left blank ...
Page 318: ...306 GC 2014 This page is intentionally left blank ...