GC-2014
1 Installation
1.1
Verification of Installation Location
Q
Installation location requirements
Consider the following points to ensure safe and appropriate unit operation when selecting
the installation site.
1.
Ambient temperature and humidity
For optimal performance, operate this unit within the following temperature and humidity
specifications.
Temperature range:
18 °C to 28 °C
Relative humidity range:
50 % to 60 % (Avoid use under conditions where
condensation forms)
Operating temperature range:
5 °C to 40 °C
Operating humidity range:
5 % to 90 % (Avoid use under conditions where con-
densation forms)
2.
Installation location
Install the unit on a firm, stable and flat base.
(The GC-2014ATF model weighs approximately 50 kg.)
3.
Corrosive gas and dust
Avoid exposure to corrosive gas and excessive dust to prolong the service life and maintain
optimal unit performance.
4.
Electro-magnetic fields and power supply noise
This unit should not be used near strong electro-magnetic fields. The power supply must
have little or no noise. These items can cause instrument problems.
5.
Other precautions
For optimal performance, avoid the following conditions during installation:
(1) Fluctuating ambient temperature.
(2) Temperature changes from heating or air conditioning.
(3) Direct sunlight.
(4) Vibrations.
Summary of Contents for GC-2014
Page 4: ...This page is intentionally left blank ...
Page 12: ...Contents viii GC 2014 This page is intentionally left blank ...
Page 28: ...1 Installation 1 3 Gas Supply Plumbing 16 GC 2014 This page is intentionally left blank ...
Page 78: ...5 Analysis 5 2 Notes for Analysis 66 GC 2014 This page is intentionally left blank ...
Page 90: ...6 Basic Key Operation 6 4 Getting Help 78 GC 2014 This page is intentionally left blank ...
Page 282: ...16 Special Functions 16 10 ROM Version No 270 GC 2014 This page is intentionally left blank ...
Page 318: ...306 GC 2014 This page is intentionally left blank ...