4 Installing Capillary Columns and Setting Analytical Flow Lines
4.1
Installation Location for Capillary Columns
Injection ports and detectors for capillary columns can be installed in positions described in
Table 4.1.1 and Fig. 4.1.1.
(Injection ports and detectors that can be used for packed columns are included.)
Table 4.1 Installation location for injection ports/detectors for capillary columns
Installation location numbers indicate the positions of column connecting joints.
Name
Possible installation location
Injection ports
SPL
1, 4
WBI
1, 4
Dual INJ
1-2, 2-3, 3-4
Single INJ
1, 4
Detectors
Dual FID
6-7
Single DET
5, 6, 7, 8
FTD-2014c
11, 12
TCD
9-10
Fig. 4.1.1
Installation location for injection ports/detectors for capillary columns