TeraScan 1.5m Polar Satellite Tracking Antenna O & M Manual
153
8.5.10 Reading PCU "N" Coefficients
The Antenna PCU contains calibration coefficients that must be reloaded when the PCU is replaced.
The software procedure described here is used to
display
the coefficients which can then be recorded so that
they can be reloaded when the PCU is replaced.
Tools:
None
Part Location:
Values are read from the
ac
command.
Procedure:
1.
The PCU must be operating and not initializing while the coefficient values are being read via the
ac
command.
2.
Enter
ac
.
3.
Enter
1
or
chain for the antenna
.
4.
Enter
n0999
. The reply will be the antenna type.
5.
Record the value displayed, using the recording format shown in step 6 for the coefficient 0.
6.
Enter the remaining six coefficient numbers (from column 3 of Table 8-1 on the following page) as in
step 4 (i.e., n1999 ENTER, n2999 ENTER, n3999 ENTER, . . .) and record the value for each as it is
displayed:
7.
Enter
reset
to reinitialize the antenna.
0: ______
4: ______
1: ______
5: ______
2: ______
6: ______
3: ______
Summary of Contents for TeraScan 1.5M
Page 2: ......
Page 6: ...iv SeaSpace Corporation...
Page 8: ...vi SeaSpace Corporation...
Page 16: ...xiv SeaSpace Corporation...
Page 26: ...10 SeaSpace Corporation...
Page 44: ...28 SeaSpace Corporation RCVCR DCVTR 3 2 9 1 5m ANTENNA 70909002 120 15...
Page 50: ...34 SeaSpace Corporation...
Page 78: ...62 SeaSpace Corporation...
Page 80: ...64 SeaSpace Corporation...
Page 82: ...66 SeaSpace Corporation...
Page 84: ...68 SeaSpace Corporation...
Page 88: ...72 SeaSpace Corporation...
Page 90: ...74 SeaSpace Corporation CBL LL 50...
Page 92: ...76 SeaSpace Corporation...
Page 128: ...112 SeaSpace Corporation...
Page 180: ...164 SeaSpace Corporation...
Page 190: ...174 SeaSpace Corporation...
Page 218: ...202 SeaSpace Corporation...
Page 226: ...210 SeaSpace Corporation...
Page 228: ...212 SeaSpace Corporation...
Page 232: ...216 SeaSpace Corporation...