ESR time (min.)
Actual duration of the ESR.
Dilution %
The calculated dilution rate after aspiration of the sample.
Temperature (°C)
Room temperature at the time of the measurement of the sample.
Pipet number
Pipette in which the sample was measured.
Error code
Shows any ESR error or warning code (e.g. "Too many borders found").
A complete overview of measurement data is shown with activating the option "Display raw data".
5.3.3 - ESR Statistics screens
A statistical graph is produced over a selected period. Make a selection of the following graphs;
▪
Daily mean (mm)
Use this to check variations in the daily mean ESR.
▪
Hazy (%)
Increasing hazy aspects are an indication for contamination of the instrument, see
▪
Dilution (%)
Increasing dilution errors indicate the need for maintenance of the diluter system.
▪
Bubbles on top (%)
Increasing samples with bubbles indicate the need for maintenance of the aspiration system, see
▪
ESR errors (%)
Increasing ESR errors may indicate the need for maintenance, see
▪
Number of samples
This can be used to document variations in work load.
Starrsed TL program
MRN-164_2-EN Version 1.08 Starrsed TL
33