Work instruction Number
242
Page
1 of 1
Purpose:
Replace O-ring inner needle
Safety:
Bio Hazard area
Instrument
: Interrliner
Revision:
001, may 2016
For assembling and disassembling the special needle holder tool TLAX110023 and spanner TLAX110026 are
needed.
Needle removal
1. Remove drip tray (metal sheet under needle unit, behind door)
2. Disconnect both tubes.
3. Loosen the knurled knob.
4. Remove the complete needle assembly outside the instrument.
5. Clean the needle with a disinfectant.
O-ring replacement
Work instruction Interrliner
118
MRN-164_2-EN Version 1.08 Starrsed TL