Work instruction Number
240
Page
1 of
1
Purpose:
Cleaning diluent system
Safety:
Bio Hazard area
Instrument
: Interrliner
Revision:
001, 2015
Step 1
1. Remove the suction-tube from the diluent pinch valve. Disconnect the tube and connect a new tube
2. Fill a bottle with 50 ml bleach (sodium hypochlorite) and 950 ml de-ionized water. (5% solution)
3. Place the suction tube in chlorine solution.
4. Use the
[Prime]
function to fill the dispenser system with the disinfectant.
5. After the prime sequence stops, press
[Prime]
to fill the dispenser system with the disinfectant.
6. Leave the disinfectant in the system for 15 minutes.
Step 2
1. Take the diluent suction tube out of the disinfectant.
2. Wipe the tube clean and dry with a tissue.
3. Empty the bottle and refill it with hot de-ionized water (80°C).
4. Place diluent suction tube in the bottle with hot water.
5. Use the
[Prime]
function.
6. After the prime sequence stops, press "
Prime
” again to fill the dispenser system with the hot water.
Step 3
1. Empty the bottle.
2. Clean the bottle with new hot de-ionized water (80°C)
3. Reconnect the diluent container.
4. Perform another
[Prime]
.
5. After the prime sequence stops press
[Prime]
again to fill the dispenser system with the new diluent solution.
Work instruction Interrliner
MRN-164_2-EN Version 1.08 Starrsed TL
117
Summary of Contents for STARRSED TL
Page 1: ...Instructions for Use Starrsed TL MRN 164_2 EN Version 1 08...
Page 2: ......
Page 7: ...Index 136 Contents MRN 164_2 EN Version 1 08 Starrsed TL 7...
Page 116: ...Work instruction Interrliner 116 MRN 164_2 EN Version 1 08 Starrsed TL...
Page 120: ...Work instruction Interrliner 120 MRN 164_2 EN Version 1 08 Starrsed TL...